Siemens IMMULITE TBG Test Kits Recalled for Inaccurate Results
Siemens Healthcare Diagnostics is recalling 155 IMMULITE TBG test kits because they may produce falsely low thyroid test results.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device that may produce inaccurate diagnostic results (depressed TBG values), which poses a risk of harm through misdiagnosis or improper treatment, but no specific injuries or illnesses have been reported in the source text.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling 155 kits of the IMMULITE/IMMULITE 1000 Systems TBG (catalog LKTB1). The affected product is used for the quantitative measurement of Thyroxine-Binding Globulin (TBG) in serum to aid in the diagnosis and treatment of thyroid diseases.
The recall was initiated because the specific lot of the product is failing the adjustment slope instrument parameter. The resulting slope of the kit is greater than 1.80, which creates the potential to produce depressed TBG patient values. This means the test may report lower levels of TBG than are actually present in the patient's serum.
The affected units are identified by LOT 0257 and SMN 10381643. These kits were distributed worldwide to Denmark, Spain, Great Britain, Germany, Bosnia Herzeg, China, and the Russian Federation. Healthcare facilities and laboratories should stop using the affected kits and contact the manufacturer for instructions on return or replacement.
The recalled product
- Product
- IMMULITE/IMMULITE 1000 Systems TBG, catalog LKTB1 Quantitative measurement of Thyroxine-Binding Globulin (TBG) in serum, as an aid in the diagnosis and treatment of thyroid diseases.
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Hazard
- Affected units
- 155
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (2)
- LOT 0257
- SMN 10381643
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- ModerateSiemens Tricyclic Antidepressant Calibrator/Controls Recall Lot U2
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · inaccurate results
See all →- HighHealthmark Protein Test Recalled Due to Expired Ingredient
FDA (Devices) · 2026-09-02
- HighHardyDisk AST Cefiderocol 30µg for In Vitro Testing Recalled Due to Incorrect Cartridges
FDA (Devices) · 2026-08-12
- HighGEM Premier 7000 with iQM3 Blood Analysis Systems Recalled for Solution Concentration Issues
FDA (Devices) · 2026-08-12
- Highepoc BGEM BUN Test Card for epoc Blood Analysis System
FDA (Devices) · 2026-05-06
- HighAbbott i-STAT EG7+ Blood Gas Cartridges Reporting Inaccurate Results
FDA (Devices) · 2026-05-06