B. Braun Outlook ES Infusion Pumps Recalled for Key Panel Failure
B. Braun Medical Inc. is recalling Outlook ES Safety Infusion Systems because the HOLD button may fail to function due to intermittent circuitry contact.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk-of-harm product (infusion pump) where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
B. Braun Medical Inc. is recalling the Outlook ES Safety Infusion System, a medical device used to regulate the flow of primary and secondary fluids when positive pressure is required. The recall involves 28,427 units distributed nationwide. The affected models are 621-100ES, 621-200ES, 621-300ES, and 621-400ES.
The recall is due to a potential issue where the key panel may fail to function as intended if the underlying circuitry fails due to intermittent contact. This issue specifically affects the HOLD button. If this failure occurs, the Door Processor (DP) will not receive the desired input from the user.
Healthcare facilities and users of these infusion pumps should contact B. Braun Medical Inc. for further instructions on how to address the issue with the key panel.
The recalled product
- Product
- Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid co
- Manufacturer
- B Braun Medical, Inc.
- Category
- Medical Device — Infusion Pump
- Affected units
- 28,427
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- Multiple serial numbers affected. Model number: 621-100ES
- 621-200ES
- 621-300ES
- and 621-400ES.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · B Braun Medical, Inc.
See all →- SevereB. Braun Recalls Lactated Ringer's Injection Due to Particulate Matter
FDA (Drugs) · 2026-09-02
- ModerateB. Braun UPMC double catheter trays contain Mastisol with cracking tubing
FDA (Devices) · 2026-06-10
- ModerateB. Braun Shands kits contain Mastisol adhesive with cracking tubing
FDA (Devices) · 2026-06-10
- ModerateB. Braun spinal epidural kits contain Mastisol with cracking tubing
FDA (Devices) · 2026-06-10
- ModerateB. Braun kits contain Mastisol adhesive with potentially cracking tubing
FDA (Devices) · 2026-06-10
Same hazard · electrical failure
See all →- HighHalcore Recalls 2025-2026 Horton Ambulances Due to Touchscreen Failure
NHTSA · 2026-08-14
- HighMarion Recalls 2025 Pumper Fire Engines Due to Display Unit Failure
NHTSA · 2026-08-10
- HighRosenbauer Recalls Emergency Vehicles with Faulty Display Units
NHTSA · 2026-08-03
- SevereBMW S 1000 RR Motorcycles Recalled Due to Potential Engine Stall Risk
NHTSA · 2026-07-23
- High