The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2143-2020·Announced 2020-06-03

Spectranetics Bridge Occlusion Balloon Recalled for Thrombus Formation Risk

Spectranetics is recalling Bridge Occlusion Balloons because current labeling is insufficient to warn users about the risk of thrombus formation if the device remains in the vasculature for long durations.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a high-risk medical device where a hazard (thrombus/embolization) is identified, but the source text does not report any specific illnesses or injuries, fitting the 'High' criteria for risk-of-harm products without reported injury.

Plain-English summary

Spectranetics Corporation is recalling Bridge Occlusion Balloons (REF 590-001) due to a potential risk of thrombus formation. The recall is driven by a determination that current labeling is insufficient to ensure that all potential users minimize the time the balloon remains in the vasculature.

The source text notes that users performing lead extraction cases may not be aware of this specific risk. If the balloon is left in the vasculature for too long, it could lead to embolization. The recall covers all lots of the product.

The product was distributed nationwide in the United States and in numerous international locations, including Australia, Canada, and various European countries. The source text does not specify specific actions for consumers or healthcare providers beyond the context of the labeling deficiency.

The recalled product

Product
Bridge Occlusion Balloon, REF 590-001, QTY 1, Sheath Compatibility 12F, Guidewire Compatibility 0.035 in, Working Length 90 cm, Balloon Outer Diameter 31 mm, Balloon Length 80 mmm, Rx ONLY, CE Marking, Sterile EO, UDI: 00813132024666

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All lots

Distribution

Distributed nationwide across the United States.

Similar recalls