The Recall Desk
HighFDA (Devices)·Z-2142-2015·Announced 2015-07-29

DePuy ATTUNE Knee Trial Instrument Recall Due to Balseal Detachment Risk

DePuy Orthopaedics is recalling ATTUNE Knee Tibial Articulation Surface Instruments because a small wire spring coil (Balseal) may detach and be left in the patient during surgery.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a risk-of-harm product (foreign object retention in a patient) where injury has not yet been reported, which aligns with the 'High' severity criteria for theoretical hazards without confirmed harm.

Plain-English summary

DePuy Orthopaedics, Inc. is recalling the ATTUNE CONV FB PS ARTICULATION SURFACE SZ1-10 instruments. These devices are used as Insert Trials during total knee replacement surgeries. The recall involves 36,607 units distributed worldwide, including the United States, Canada, and numerous international locations.

The recall is due to a potential defect where the Balseal, a small wire spring coil located on the post feature of the instrument, may become damaged and disassociate (come off). If the Balseal separates from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

The source text does not specify specific actions for consumers or surgeons, nor does it report any specific injuries or illnesses. The hazard is identified as a risk of foreign object retention during the surgical trialing process.

The recalled product

Product
ATTUNE CONV FB PS ARTICULATION SURFACE SZ1-10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation
Affected units
36,607

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (19)

  • Product Codes:254500971
  • 254500972
  • 254500973
  • 254500974
  • 254500975
  • 254500976
  • 254500977
  • 254500978
  • 254500979
  • and 254500980. GTIN: 10603295135487
  • 10603295135494
  • 10603295135500
  • 10603295135517
  • 10603295135524
  • 10603295135531
  • 10603295135548
  • 10603295135555
  • 10603295135562
  • and 10603295135579.

Distribution

Part of a larger recall action

This product is one of 10 recalled by DePuy Orthopaedics, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 10 products in this recall →