The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2099-2021·Announced 2021-07-28

Medical Device Calibration Standard Assigned Incorrectly in NEFA CAL Kit

Randox Laboratories recalls NEFA CAL IVD calibration standard kits (Batch 544642) due to incorrectly assigned calibration values that could cause measurement errors up to 8% in quality control and patient test results.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: FDA Class II medical device—a risk-of-harm product where injury has not yet been reported. The incorrectly assigned calibration standard would cause systematic measurement errors up to 8% in clinical laboratory results, potentially affecting diagnostic accuracy and patient care decisions.

Plain-English summary

Randox Laboratories Ltd. is recalling 71 kits of NEFA CAL (Non-Esterified Fatty Acids) IVD Standard for measurement, Catalog Number FA115, Batch Number 544642, with expiration date November 28, 2022.

The calibration standard value has been assigned incorrectly. When used to calibrate analytical equipment, this incorrect standard will cause a positive shift of up to 8% in both quality control measurements and patient sample results.

The affected kits were distributed nationwide in Arkansas, California, Connecticut, New York, Ohio, Pennsylvania, Tennessee, and Texas.

No reported illnesses or adverse events related to this product have been identified. Laboratories using this batch should discontinue use and contact Randox Laboratories for replacement kits or additional guidance.

The recalled product

Product
NEFA CAL (Non-Esterified Fatty Acids)- IVD Standard for measurement of NEFA Catalog Number: FA115
Affected units
71

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Batch Number: 544642 Exp. Date: 28th Nov 2022 GTIN: 05055273203066

Distribution

Distributed nationwide across the United States.

The notice also names 8 states specifically: