The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-2093-2018·Announced 2018-06-13

Tosoh ST-AIA PACK Testosterone Assay Recalled Due to Interference

Tosoh Bioscience Inc is recalling ST-AIA PACK Testosterone assays because Asfotase Alfa interferes with the tests, potentially causing inaccurate results.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves potential diagnostic errors (falsely increased or decreased results) which poses a risk of harm to patients through incorrect medical decisions, fitting the 'High' severity criteria for risk-of-harm products.

Plain-English summary

Tosoh Bioscience Inc is recalling the ST-AIA PACK Testosterone Assay, Part Number 025204. The recall involves 10,913 units distributed worldwide, including the United States, Canada, Mexico, and several countries in South and Central America.

The recall is being issued because Asfotase Alfa (Strensiq) interferes with certain Tosoh assays. This interference can potentially cause falsely increased or falsely decreased immunoassay test results, depending on the specific assay used.

Healthcare providers and laboratories should stop using the affected product and contact Tosoh Bioscience Inc for further instructions regarding the recall and potential replacement or refund.

The recalled product

Product
ST-AIA PACK Testosterone; Part Number: 025204 Assay, Reproductive Hormone
Affected units
10,913

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All Lot Codes

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 35 recalled by Tosoh Bioscience Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 35 products in this recall →