The Recall Desk
HighFDA (Devices)·Z-2044-2013·Announced 2013-08-28

Zimmer Recalls NexGen Knee Components Due to Unapproved Combinations

Zimmer, Inc. is recalling NexGen CR Complete Knee System components because surgeons are implanting them in unapproved combinations.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves the use of medical devices in unapproved combinations, which presents a risk of harm even though specific injuries are not detailed in the source text.

Plain-English summary

Zimmer, Inc. is recalling specific components of the NexGen Cruciate Retaining (CR) Complete Knee System. The recall involves CR Micro Femurs, Articular Surface, Patellae, instruments, and provisionals that are being combined with standard components in ways that are not approved.

The recall is being issued because surgeons are implanting these components in unapproved combinations, specifically mixing micro and standard parts. A previous recall in March 2012 reiterated the proper combination labeling, but the mismatch problem has continued. The affected devices are used in total knee arthroplasty and are indicated for patients with severe pain.

The recall covers 192,355 devices distributed nationwide in the United States and in Japan. Consumers and healthcare providers should contact Zimmer, Inc. for further information regarding the removal of these devices.

The recalled product

Product
00597104110 Provisional CR ART SURF PROV 56/STR GRN 10, Rx, Sterile; 00597104112 Provisional CR ART SURF PROV 56/STR GRN 12, Rx, Sterile; 00597104114 Provisional CR ART SURF PROV 56/STR GRN 14, Rx, Sterile; 00597104117 Provisional CR ART SURF PROV 56/STR GRN 17, Rx,
Manufacturer
Zimmer, Inc.
Affected units
192,355

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • all codes

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 47 recalled by Zimmer, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 47 products in this recall →