The Recall Desk
Agency classClass IIOur severity:ModerateFDA (Devices)·Z-2035-2017·Announced 2017-05-10

Beckman Coulter AQUIOS CL Flow Cytometer Software Export Defect Recall

Beckman Coulter is recalling AQUIOS CL Flow Cytometers because the software's Export feature incorrectly handles Tetra Combo CRD files.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 2/5 (Moderate).

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, and the source text describes a software export defect without reporting any specific injuries, illnesses, or deaths, fitting the criteria for a moderate severity precautionary recall.

Plain-English summary

Beckman Coulter Inc. is recalling the AQUIOS CL Flow Cytometer, Reference number B30166, equipped with Software Versions 2.0 and 2.0.1. The recall involves 188 units total, with 24 units distributed in the United States and the remainder distributed worldwide in various countries.

The recall was initiated because the Export feature located in the Results area of the AQUIOS System software is not correctly exporting Tetra Combo CRD files. This defect affects the in vitro diagnostic flow cytometric applications for which the device is intended.

Users of the affected devices should contact Beckman Coulter for further instructions regarding the software defect.

The recalled product

Product
AQUIOS CL Flow Cytometer, Ref no. B30166, Software Versions 2.0 and 2.0.1 Product Usage: The AQUIOS CL Flow Cytometer is intended for use with in vitro diagnostic flow cytometric applications using up to four fluorescent detection channels using a blue (488 nm) laser, two…

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • All

Distribution

Distributed nationwide across the United States.