Siemens AXIOM Artis TA X-ray Generator Recall for Power Supply Issue
Siemens is recalling five AXIOM Artis TA units due to a power supply tolerance issue that can cause generator failure and prevent X-ray imaging.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the source text explicitly states that no injuries or illnesses have been reported, fitting the criteria for a moderate severity score.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling five units of the AXIOM Artis TA, Model Number 7007755. The recall is due to a tolerance issue in the power supply of the generator control that may cause the generator A100 to fail. The affected serial numbers are 32090, 32085, 32091, 32094, and 32095.
The issue occurs sporadically and may happen during an ongoing procedure. If the generator fails, the X-ray tube will not be supplied with the required voltage, making it impossible to perform X-ray imaging. The units were distributed nationwide in the United States.
This is a Class II recall. No injuries or illnesses have been reported. Healthcare facilities should contact Siemens Medical Solutions USA, Inc. for instructions on how to resolve the issue.
The recalled product
- Product
- AXIOM Artis TA, Model Number 7007755
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Category
- Medical Device — X-ray Equipment
- Hazard
- Affected units
- 5
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Serial Numbers: 32090 32085 32091 32094 32095
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 23 recalled by Siemens Medical Solutions USA, Inc under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 23 products in this recall →Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · equipment failure
See all →- HighGE Optima XR200/220/240amx Radiographic System Column Safety Mechanism
FDA (Devices) · 2026-07-22
- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22
- SevereOmnipod Eros Pod Insulin Delivery Failure Risk
FDA (Devices) · 2026-07-08
- SevereCyberknife Treatment Delivery System Collimator Docking and Fall Hazard Recall
FDA (Devices) · 2026-07-08
- HighPetzl America Recalls Three Climbing Harness Models for Fall Hazard
CPSC · 2026-05-14