The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-1836-2019·Announced 2019-06-26

Signia Tri-Staple Surgical Device Recalled Due to Missing Pin Components

The FDA is recalling Signia Tri-Staple 2.0 devices because missing pin components may cause incomplete staple formation, leading to serious surgical complications.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as FDA Class II and involves a defect that can lead to serious injuries such as sepsis, peritonitis, and pneumothorax, meeting the criteria for a Severe rating.

Plain-English summary

The U.S. Food and Drug Administration (FDA) has issued a Class II recall for the Signia Tri-Staple 2.0 Curved Tip 60mm Medium/Thick surgical stapler devices manufactured by Covidien LLC. The recall involves 3,113,280 units that were distributed nationwide and worldwide.

The recall was initiated because the devices may be missing one of two pin components that maintain the alignment of the device jaws. This defect can result in incomplete staple formation during use. Incomplete staple formation may lead to bleeding, anastomotic leak, peritonitis, or pneumothorax, which can result in the potential for infection and/or sepsis.

Healthcare providers and facilities should stop using the affected devices and contact the manufacturer for instructions on return or replacement. The specific lot codes affected are listed in the official recall notice.

The recalled product

Product
SIGNIA TRI-STAPLE 2.0 CURVED TIP 60MM MEDIUM/ THIC , SIG60CTAMT
Manufacturer
Covidien LLC
Affected units
3,113,280

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 12 recalled by Covidien LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 12 products in this recall →

Similar recalls