Recall of Covidien Signia Stapler Reloads Due to Safety Interlock Failure
FDA is recalling specific lots of Covidien Signia stapler reloads because a safety interlock may fail, allowing empty reloads to be fired.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The recall is classified as FDA Class II and involves a device malfunction that can lead to significant patient injury, including bleeding, tissue trauma, and the need for open surgery.
Plain-English summary
The U.S. Food and Drug Administration (FDA) is recalling specific lots of the Covidien Signia Small Diameter Curved Tip Intelligent Reload 30 mm Vascular/Thin 8 mm - Short (Item Code: SIGSDS30CTVT). The recall involves 15,298 units distributed worldwide, including nationwide distribution in the United States and in Austria, Belgium, Chile, China, Denmark, Finland, France, Germany, Hong Kong, Ireland, Italy, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, and the United Kingdom.
The recall is due to a potential failure of the device's safety interlock. This interlock is designed to prevent an empty single-use reload from being fired a second time. If the interlock fails and the empty reload is fired, the stapler may cut tissue without deploying staples. This malfunction can result in a failed anastomosis, bleeding, tissue trauma, and the potential need to convert to an open surgical procedure.
Healthcare providers and facilities should check their inventory for the specific lot numbers and UDI codes listed in the official recall notice. If the affected devices are in use, they should be removed from service and returned to the manufacturer or disposed of according to the recall instructions provided by Covidien, LP.
The recalled product
- Product
- Covidien Signia Small Diameter Curved Tip Intelligent Reload 30 mm Vascular/Thin 8 mm - Short Item Code:SIGSDS30CTVT
- Manufacturer
- Covidien, LP
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 3 recalled by Covidien, LP under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 3 products in this recall →Related recalls
Same manufacturer · Covidien, LP
See all →- SevereSurgipro II Sutures Recalled for Potential Sterile Barrier Defects
FDA (Devices) · 2026-08-19
- SevereDermalon Sutures Recalled for Potential Sterile Barrier Packaging Defects
FDA (Devices) · 2026-08-19
- HighV-Loc 90 Absorbable Wound Closure Devices Recalled for Packaging Defects
FDA (Devices) · 2026-08-19
- HighSurgipro Sutures Recalled for Potential Sterile Barrier Perforation
FDA (Devices) · 2026-08-19
- HighV-Loc 180 Absorbable Wound Closure Device Recalled for Packaging Defects
FDA (Devices) · 2026-08-19
Same hazard · device malfunction
See all →- SevereMedtronic MiniMed Flex Insulin Pumps Recalled for Failed Verification Tests
FDA (Devices) · 2026-10-14
- ModerateWilson-Cook Recalls Hemospray Endoscopic Hemostat Due to Adherence Issues
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Aisys CS2 Anesthesia Systems Recalled for Gas Mixer Defect
FDA (Devices) · 2026-09-23
- HighDatex-Ohmeda Et Control Software Recalled for Anesthetic Delivery Interruption
FDA (Devices) · 2026-09-23
Similar recalls
- HighCovidien Signia Surgical Stapler Reloads Recalled Due to Articulation Defect
FDA (Devices) · 2024-04-10
- HighSurgical staplers recalled over a malformed cover tube press
FDA (Devices) · 2025-11-19
- HighSignia Surgical Stapler Reload recalled for uncontrolled articulation
FDA (Devices) · 2024-04-10
- HighFDA Recalls 222 Units of Covidien Tri-Staple 2.0 Reinforced Intelligent Reloads
FDA (Devices) · 2021-04-21
- SevereRecall of Signia Tri-Staple 2.0 Surgical Reloads Due to Missing Pins
FDA (Devices) · 2019-06-26
- HighVATS lobectomy stapler boxes recalled over a malformed cover tube press
FDA (Devices) · 2025-11-19