St. Jude Medical Recalls Eon and Eon Mini Charging Systems

St. Jude Medical is recalling Eon and Eon Mini charging systems due to reported warmth or heating at the implant site during charging.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device with a risk of harm (heating at the implant site) where the source text does not explicitly report injuries or illnesses, fitting the criteria for a High severity score.

Plain-English summary

St. Jude Medical is recalling the Eon and Eon Mini charging systems, including model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger), and Eon Mini -3721 (Charger). This recall is a continuation of the firm's July 2012 recall and involves 75,663 units distributed worldwide, including the United States and numerous international countries.

The recall is being conducted because of issues involving warmth or heating at the implant site during charging for the Eon Mini implantable pulse generators (IPGs). The source text indicates that the firm previously informed customers of design improvements to address possible increased energy dissipation when the charger is misaligned with the IPG or the IPG is implanted too near the surface.

The affected products are used to recharge IPG batteries by transmitting RF energy through an antenna. The Eon Wall Charger and Portable Charger allow for recharging, with the portable version allowing patient mobility. The Eon Mini Charger allows recharging while stimulation is either on or off. Consumers and healthcare providers should refer to the recalling firm for specific instructions on handling these devices.

The recalled product

Product
Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 recall. Product Usage: The Eon Wall Charger (3701) provides the capability to recharge the IPG…
Manufacturer
St. Jude Medical
Affected units
75,663