GXL Acetabular Liners Packaging Defect May Cause Accelerated Wear
Exactech GXL acetabular hip implant liners in non-conforming bags may experience accelerated wear. Approximately 7,946 devices were distributed worldwide. No injuries have been reported.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This Class II recall addresses a risk-of-harm product (orthopedic joint implant) with no reported injuries. The non-conforming packaging causing accelerated wear in a critical structural component meets the rubric criterion for Score 3: a risk-of-harm product where injury has not yet been reported.
Plain-English summary
Exactech, Inc. is recalling certain GXL acetabular polyethylene liners used in hip replacement surgery. These orthopedic implant components were packaged in bags that do not conform to specifications.
The non-conforming bags may adversely impact the liners and contribute to accelerated wear of the device. Because acetabular liners are critical load-bearing components in hip implants, accelerated wear could potentially affect implant performance and longevity.
Approximately 7,946 affected devices were distributed worldwide. In the United States, distribution was nationwide. Outside the US, the liners were distributed to Argentina, Brazil, Chile, China, Colombia, Ecuador, Greece, Guatemala, India, Lebanon, Portugal, Puerto Rico, and Spain.
Patients who received hip implants containing GXL liners should contact their surgeon or healthcare provider with questions about their specific implant. Healthcare providers can identify affected devices by checking serial numbers and reference codes provided in the recall notice.
The recalled product
- Product
- GXL acetabular liners, orthopedic hip implant component. Product Line Numbers: a. ACUMATCH 0 DEGREE LINER 28MM 46MM, REF 130-28-05; b. ACUMATCH 0 DEGREE LINER 28MM 48MM, REF 130-28-06; c. ACUMATCH 0 DEGREE LINER 28MM 50/52MM, REF 130-28-07; d. ACUMATCH 0 DEGREE LINER…
- Manufacturer
- Exactech, Inc.
- Affected units
- 7,946
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- a. REF 130-28-05
- GUDID 10885862021915
- Serial Numbers: 693079
- 693081
- 693083
- 693091
- 793963
- 793968
- 793970
- 793982
- b. REF 130-28-06
- GUDID 10885862021922
- Serial Numbers: 618681
- 618682
- 618683
- 618686
- 618688
- 618691
- 618693
- 618696
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 10 recalled by Exactech, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 10 products in this recall →Related recalls
Same manufacturer · Exactech, Inc.
See all →- HighExactech Ergo modular impactor handle may be missing its cross-pin
FDA (Devices) · 2026-03-18
- HighExactech Equinoxe core instrument kits may have handles missing cross-pins
FDA (Devices) · 2026-03-18
- ModeratePilot tip converters recalled after release without completed performance testing
FDA (Devices) · 2025-11-26
- ModerateShoulder reamers recalled after release without completed performance testing
FDA (Devices) · 2025-11-26
- ModerateShoulder reamer 46mm recalled after release without completed performance testing
FDA (Devices) · 2025-11-26
Same hazard · accelerated wear
See all →- HighGXL Acetabular Liners Recalled Due to Non-Conforming Packaging Bags
FDA (Devices) · 2022-09-21
- HighExactech GXL Acetabular Hip Implant Liners Recalled for Accelerated Wear
FDA (Devices) · 2022-09-21
- HighGXL Acetabular Hip Implant Liners Recalled Due to Packaging Defect
FDA (Devices) · 2022-09-21
- HighGXL Acetabular Liners Recalled Due to Non-Conforming Packaging
FDA (Devices) · 2022-09-21
- ModerateGXL Acetabular Liners Recalled Due to Non-Conforming Packaging
FDA (Devices) · 2022-09-21
Similar recalls
- HighExactech Connexion GXL hip implant liners risk edge-loading and wear
FDA (Devices) · 2021-07-28
- HighExactech Connexion GXL hip implant liners recalled due to premature wear risk
FDA (Devices) · 2021-07-28
- HighHip implant liners recalled for edge-loading and premature wear
FDA (Devices) · 2021-07-28
- HighExactech Hip Implant Liners Recalled Due to Premature Wear Risk
FDA (Devices) · 2021-07-28
- HighExactech Connexion GXL Hip Implant Liners Recalled for Premature Wear Risk
FDA (Devices) · 2021-07-28
- HighHip Implant Acetabular Liners Recalled for Premature Wear Risk
FDA (Devices) · 2021-07-28