The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1724-2013·Announced 2013-07-17

Biomet Recalls Oxford Fixed Bearing Partial Knee Components

Biomet is recalling 24 Oxford Fixed Bearing Partial Knee components due to a design flaw that may cause bone impingement.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device with a design defect that creates a potential for injury (impingement), but the source text does not report any actual injuries or illnesses, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.

Plain-English summary

Biomet, Inc. is recalling 24 units of the REF 154239 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia, Size F7. These medical devices are partial knee replacement systems intended for use in the lateral compartment of the knee. The affected units were distributed worldwide, including in the United States (North Carolina, Indiana, Pennsylvania, and Ohio), Canada, and the Netherlands.

The recall was initiated after a Biomet investigation found that the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension. This design issue poses a risk of harm to patients who have had these specific components implanted.

Healthcare providers and patients should contact Biomet for instructions on how to proceed. The source text does not specify if any injuries or illnesses have been reported to date.

The recalled product

Product
REF 154239 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size F7 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing…
Manufacturer
Biomet, Inc.
Affected units
24

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (5)

  • M492980
  • M657130
  • M920670
  • M886460
  • M166900

Distribution

Distributed in 4 states:

Part of a larger recall action

This product is one of 30 recalled by Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 30 products in this recall →

Similar recalls