RayStation Therapy Treatment Planning Software Recalled for Synchronization Bug
RaySearch Laboratories is recalling RayStation Therapy Treatment Planning System software due to a bug that can cause views to become unsynchronized with stored data.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 2/5 (Moderate).
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II. The source text explicitly states that the bug does not affect dose computations, indicating a theoretical risk of data synchronization issues rather than a direct risk of serious injury or death, which aligns with a Moderate severity rating for a software defect without reported harm.
Plain-English summary
RaySearch Laboratories AB is recalling RayStation Therapy Treatment Planning System Stand-alone Software versions 3.0, 3.5, 4.0, 4.5, and 4.7. The recall involves 1,264 total units, with 552 units distributed domestically and 711 units distributed internationally.
The recall is due to a software issue with editing tools that use the left mouse button held down for drawing and interacting with patient views and beams eye views. The views can become unsynchronized with the stored data if a user simultaneously right-clicks, presses Ctrl-S, or presses Ctrl-Z while holding the left mouse button. The source text states that this bug does not affect dose computations, which are based on the stored system state.
The affected software was distributed in the United States (AR, AZ, CA, CT, CO, DE, FL, GA, IL, LA, MI, MD, MS, MA, NY, NV, NJ, OH, OR, PA, NC, RI, TN, WA, and Washington D.C.) and internationally in Algeria, Australia, Austria, Belgium, Canada, China, France, French Polynesia, Germany, Iran, Italy, Japan, Netherlands, New Zealand, Norway, Republic of Korea, Spain, Sweden, Switzerland, Thailand, and the United Kingdom. Users should contact RaySearch Laboratories AB for further instructions regarding the software update or resolution.
The recalled product
- Product
- RayStation Therapy Treatment Planning System Stand-alone Software 3.0, 3.5, 4.0, 4.5 and 4.7., designed for treatment planning and analysis of radiation therapy.
- Manufacturer
- RAYSEARCH LABORATORIES AB
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (14)
- 3.0.0.251
- 3.5.0.16
- 3.5.1.6
- 4.0.0.14
- 4.0.1.4
- 4.0.2.9
- 4.0.3.4
- 4.5.0.19
- 4.5.1.14
- 4.5.2.7
- 4.7.0.15
- 4.7.1.10
- 4.7.2.5
- 4.7.3.13 or 4.7.4.4
Distribution
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · software bug
See all →- HighMerge PACS Medical Imaging Workstation Recalled Due to Patient Data Display Error
FDA (Devices) · 2026-08-12
- HighMiniMed Go App Recalled Due to Missed Insulin Alerts
FDA (Devices) · 2026-07-29
- HighImaging viewer software recalled over inaccurate length measurement values
FDA (Devices) · 2025-12-24
- HighImaging software recalled over a bug that can lose clinical information
FDA (Devices) · 2025-12-03
- HighEPIONE Stereotactic Device Software Bug Poses Instrument Guidance Risk
FDA (Devices) · 2025-06-18
Similar recalls
- HighRayStation Radiation Therapy Software Recalled for Potential Dose Errors
FDA (Devices) · 2014-04-09
- HighRayStation Radiation Therapy Software Recalled for Beam Editing Bug
FDA (Devices) · 2020-01-01
- HighRayStation Software Recall for Radiation Therapy Dose Calculation Errors
FDA (Devices) · 2014-06-25
- ModerateRaySearch Radiation Therapy Treatment Planning System Software Recall
FDA (Devices) · 2016-10-19
- SevereRaySearch Ray Station 4.0 software recalled for dose calculation errors
FDA (Devices) · 2014-04-09
- HighRayStation software recall for potential incorrect treatment plan references
FDA (Devices) · 2020-12-02