The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1710-2013·Announced 2013-07-17

Biomet Recalls Oxford Fixed Bearing Partial Knee Components

Biomet is recalling 29 Oxford Fixed Bearing Partial Knee components due to a design defect that may cause anterior impingement.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a design defect creating a potential for impingement, which is a risk-of-harm product where specific injury reports are not detailed in the source text, fitting the 'High' severity criteria for theoretical or unreported harm risks.

Plain-English summary

Biomet, Inc. is recalling 29 units of the Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia (REF 154221, Size D3). These medical devices are intended for use in the lateral compartment of the knee and are implemented with bone cement.

The recall was initiated after a Biomet investigation found that the design of the tibial bearing creates the potential for anterior impingement between the bone and the poly bearing component in terminal extension.

The affected units were distributed worldwide, including in the United States (North Carolina, Indiana, Pennsylvania, and Ohio), Canada, and the Netherlands. Healthcare providers should stop using these devices and contact Biomet for instructions on return or replacement.

The recalled product

Product
REF 154221 Oxford Fixed Bearing Partial Knee Cemented Lateral Tibia Size D3 ARCOM UHMWPE / CO-CR-MO Alloy Only for use in Lateral compartment Sterile R The Oxford Fixed Bearing Knee is a Partial knee replacement system consisting of a femoral component and a fixed bearing…
Manufacturer
Biomet, Inc.
Affected units
29

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (6)

  • M987290
  • M035670
  • 987410
  • M444830
  • M920680
  • M754840

Distribution

Distributed in 4 states:

Part of a larger recall action

This product is one of 30 recalled by Biomet, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 30 products in this recall →