T25 Cannulated Screwdrivers May Break During Surgical Procedures
T25 Cannulated Screwdrivers (CSD 025) packaged in Guide Wire Kits may break at the driver tip during use, extending surgery time and requiring additional imaging to identify fragments.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II recall of a surgical device that may break during use, creating a risk of harm through extended surgery and potential fragment retention. No injuries or hospitalizations have been reported; the hazard is risk-based rather than incident-based.
Plain-English summary
Flower Orthopedics Corporation is recalling T25 Cannulated Screwdrivers (CSD 025) packaged in Guide Wire Kits (GWK 300). These screwdrivers are designed for inserting orthopedic screws with T25 driving features in diameters of 5.5mm, 6.5mm, and 7.3mm.
The screwdrivers may break at the driver tip along the shaft near the handle during use. If breakage occurs, fragments may remain in the surgical site, extending surgery time and requiring additional x-ray imaging to locate and identify the fragments.
This recall affects 47 units distributed nationwide to healthcare facilities in Arizona, Colorado, Florida, Georgia, Illinois, Indiana, Pennsylvania, North Carolina, and Virginia. The affected lot numbers are 2003320898, 1810220041, 1802220008, 1802220005, and 1711220035 (UDI: 00840118110853).
If you have received this product, verify the lot number on your inventory. Do not use the affected screwdrivers and contact Flower Orthopedics Corporation with any questions about the recall.
The recalled product
- Product
- T25 Cannulated Screwdrivers (CSD 025) packaged in Guide Wire Kits (GWK 300) - Product Usage: Intended to be used to insert screws with T25 driving features, which includes screws with diameters of 5.5mm, 6.5mm, and 7.3mm Product Code: GWK 300
- Manufacturer
- Flower Orthopedics Corporation
- Affected units
- 47
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- Lot Numbers: 2003320898
- 1810220041
- 1802220008
- 1802220005
- 1711220035 UDI: 00840118110853
Distribution
Related recalls
Same manufacturer · Flower Orthopedics Corporation
See all →- HighFlower Orthopedics Cannulated Reamer Recalled for Breaking During Surgical Use
FDA (Devices) · 2023-10-18
- HighOrthopedic Inserter Shaft Recalled for Mating Incompatibility During Surgery
FDA (Devices) · 2022-01-12
- HighFlower Drill Bit Kit 2.4mm missing AO connect feature
FDA (Devices) · 2021-10-20
- ModerateFlower Drill Bit Kit Missing AO Connect Feature, May Delay Procedures
FDA (Devices) · 2021-10-20
- ModerateFlower Drill Bit Kit missing AO connect feature may delay procedures
FDA (Devices) · 2021-10-20
Same hazard · device breakage
See all →- HighTissium Peripheral Nerve Repair Device Shelf Life Recall
FDA (Devices) · 2026-08-12
- HighPeripheral Nerve Repair Device Relabeled Due to Unsubstantiated Shelf Life
FDA (Devices) · 2026-08-12
- HighTissium COAPTIUM CONNECT Nerve Repair Device Shelf Life Recall
FDA (Devices) · 2026-08-12
- HighSurgify Halo Surgical Burr Breakage Risk During Spinal Surgery
FDA (Devices) · 2026-07-01
- SevereSurgify Halo 4.0mm Extendable Surgical Burr Recall for Breakage Risk
FDA (Devices) · 2026-07-01
Similar recalls
- SevereFlower Orthopedics Recalls Cannulated Drill Bits Due to Breakage During Surgery
FDA (Devices) · 2017-02-08
- SevereIntegra Hallu-Fix Cannulated Screw Driver Recall Due to Tip Breakage
FDA (Devices) · 2013-07-17
- ModerateTrilliant Surgical Recalls 3.0/4.0 Cannulated Drill Bits Due to Sizing Issue
FDA (Devices) · 2015-09-02
- HighFlower Orthopedics Cannulated Reamer Recalled for Breaking During Surgical Use
FDA (Devices) · 2023-10-18
- SevereTrilliant Gridlock Ankle Screw Driver Bit Recalled for Specification Defect
FDA (Devices) · 2020-01-22
- HighStraumann SCS Screwdriver Recalled for Incorrect Tip Configuration
FDA (Devices) · 2021-08-11