Maquet Getinge Children's ECC Tubing Sets Recalled for Sterility Assurance Concerns
Maquet Cardiovascular is recalling specific Children's Large ECC Packs because a design feature may prevent proper sterilization, potentially compromising sterility.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device where sterility cannot be assured, posing a risk of infection or harm, but no specific injuries or illnesses are reported in the source text.
Plain-English summary
Maquet Cardiovascular, LLC is recalling the Maquet Getinge-BEQ-TOP 49101 Childrens Large ECC Pack (Material: 701067069R01). The recall affects units with Batch Number 3000059724 that were distributed nationwide.
The recall was initiated because the closed stopcock and non-vented cap configuration on the custom tubing sets may prevent the flow of humidity and ethylene oxide gas into the fluid path during sterilization. As a result, the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured, meaning the devices may not be sterile.
Healthcare facilities and consumers should stop using these specific tubing sets immediately. Users should contact Maquet Cardiovascular for instructions on how to return or dispose of the affected products.
The recalled product
- Product
- Maquet Getinge-BEQ-TOP 49101 Childrens Large ECC Pack Material: 701067069R01
- Manufacturer
- Maquet Cardiovascular, LLC
- Category
- Medical Device — Cardiovascular
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Batch Numbers: 3000059724
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 183 recalled by Maquet Cardiovascular, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 183 products in this recall →Related recalls
Same manufacturer · Maquet Cardiovascular, LLC
See all →- HighVasoview Hemopro 2 Endoscopic Vessel Harvesting System Recall
FDA (Devices) · 2026-07-22
- HighVessel harvesting systems recalled after out-of-box failures in harvesting tool jaws
FDA (Devices) · 2025-10-01
- HighHeartstring III seal systems recalled over failures to load, deploy and seal
FDA (Devices) · 2025-09-24
- HighHeartstring III 4.3 mm seal systems recalled over loading and deployment failures
FDA (Devices) · 2025-09-24
- HighHeartstring III 3.8 mm seal systems recalled over three identified failure modes
FDA (Devices) · 2025-09-24
Same hazard · sterility failure
See all →- SevereAvKare Carboxymethylcellulose Sodium Eye Drops Recalled for Sterility Concerns
FDA (Drugs) · 2026-09-30
- SevereAvKare Carboxymethylcellulose Sodium Eye Drops Recalled for Sterility Concerns
FDA (Drugs) · 2026-09-30
- SevereROHTO DIGI-EYE Eye Drops Recalled Over Sterility Assurance Gap
FDA (Drugs) · 2026-07-29
- HighROHTO DRY-AID Eye Drops Recalled Due to Sterility Concerns
FDA (Drugs) · 2026-07-29
- SevereOne-Piece Introducer Tray kits recalled for sterility failure
FDA (Devices) · 2026-07-22
Similar recalls
- HighMaquet CARDIOHELP System extracorporeal support device components recalled for sterility
FDA (Devices) · 2023-03-08
- HighMaquet Custom Tubing Packs Recalled for Potential Endotoxin Exposure
FDA (Devices) · 2016-04-20
- SevereDatascope Recalls Better Bladder Devices in Custom Tubing Kits
FDA (Devices) · 2018-05-16
- HighDatascope Recalls Pediatric Cardiopulmonary Bypass Tubing Kits Due to Valve Defect
FDA (Devices) · 2018-06-20
- HighGetinge Pediatric Diffusion Membrane Oxygenator Recalled for Sterility Concerns
FDA (Devices) · 2019-12-18
- HighMaquet Recalls HLS Set Advanced Oxygenator Tubing Packs Due to Endotoxin Risk
FDA (Devices) · 2016-04-20