The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1631-2019·Announced 2019-06-05

Maquet Getinge ECC Tubing Sets Recalled for Sterility Assurance Concerns

Maquet Cardiovascular is recalling specific ECC tubing sets because a design flaw may prevent proper sterilization, potentially compromising sterility.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device where sterility cannot be assured, posing a risk of infection (risk-of-harm product) without reported injuries, which aligns with the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.

Plain-English summary

Maquet Cardiovascular, LLC is recalling Maquet Getinge-BEQ-TOP 17207 3/8" ECC Packs (Material: 701049346R01) due to a potential sterility issue. The recall affects Batch Number 3000079611, which was distributed nationwide.

The recall was initiated because the closed stopcock and non-vented cap configuration may prevent the flow of humidity and ethylene oxide gas into the fluid path during sterilization. As a result, the Sterility Assurance Level (SAL) for the stopcock ports cannot be assured, meaning the devices may not be sterile when used.

Healthcare facilities and users of these specific tubing sets should stop using the affected units and contact Maquet Cardiovascular for further instructions on returns or replacements.

The recalled product

Product
Maquet Getinge-BEQ-TOP 17207 3/8" ECC Pack Material:701049346R01

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Batch Numbers: 3000079611

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 183 recalled by Maquet Cardiovascular, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 183 products in this recall →