Maquet Getinge-BEQ-TOP 5208 ECC Set Cardiac Recall for Sterility Assurance
Maquet Cardiovascular is recalling specific batches of cardiac extracorporeal circulation sets because a design flaw may prevent proper sterilization of the stopcock ports.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device where the Sterility Assurance Level cannot be assured, representing a risk of harm (infection) where injury has not yet been reported, fitting the 'High' severity criteria for risk-of-harm products.
Plain-English summary
Maquet Cardiovascular, LLC is recalling the Maquet Getinge-BEQ-TOP 5208 ECC SET CARDIAC (Material: 701067350). The recall affects specific batch numbers: 3000044413, 3000049064, 3000050885, 3000054157, 3000058832, 3000062461, and 3000065084. The product was distributed nationwide.
The recall was initiated because the closed stopcock and non-vented cap configuration may prevent the flow of humidity and ethylene oxide gas into the fluid path during sterilization. As a result, the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured, creating a potential risk of non-sterile components being used in extracorporeal circulation procedures.
Healthcare facilities and users should identify any affected units by checking the batch numbers listed above. Users should stop using the affected products and contact Maquet Cardiovascular, LLC for further instructions on return or replacement.
The recalled product
- Product
- Maquet Getinge-BEQ-TOP 5208 ECC SET CARDIAC Material: 701067350
- Manufacturer
- Maquet Cardiovascular, LLC
- Category
- Medical Device — Cardiac Surgery
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 183 recalled by Maquet Cardiovascular, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 183 products in this recall →Related recalls
Same manufacturer · Maquet Cardiovascular, LLC
See all →- HighVasoview Hemopro 2 Endoscopic Vessel Harvesting System Recall
FDA (Devices) · 2026-07-22
- HighVessel harvesting systems recalled after out-of-box failures in harvesting tool jaws
FDA (Devices) · 2025-10-01
- HighHeartstring III seal systems recalled over failures to load, deploy and seal
FDA (Devices) · 2025-09-24
- HighHeartstring III 4.3 mm seal systems recalled over loading and deployment failures
FDA (Devices) · 2025-09-24
- HighHeartstring III 3.8 mm seal systems recalled over three identified failure modes
FDA (Devices) · 2025-09-24
Same hazard · sterility failure
See all →- SevereAvKare Carboxymethylcellulose Sodium Eye Drops Recalled for Sterility Concerns
FDA (Drugs) · 2026-09-30
- SevereAvKare Carboxymethylcellulose Sodium Eye Drops Recalled for Sterility Concerns
FDA (Drugs) · 2026-09-30
- SevereROHTO DIGI-EYE Eye Drops Recalled Over Sterility Assurance Gap
FDA (Drugs) · 2026-07-29
- HighROHTO DRY-AID Eye Drops Recalled Due to Sterility Concerns
FDA (Drugs) · 2026-07-29
- SevereOne-Piece Introducer Tray kits recalled for sterility failure
FDA (Devices) · 2026-07-22
Similar recalls
- HighMaquet CARDIOHELP System extracorporeal support device components recalled for sterility
FDA (Devices) · 2023-03-08
- HighMaquet Cardiovascular Recalls HLS Set Advanced Due to Sterility Breach
FDA (Devices) · 2020-12-23
- HighMaquet Cardiovascular Recalls HLS Set Advanced Due to Sterility Breaches
FDA (Devices) · 2020-12-23
- LowDatascope Cardiopulmonary Bypass Tubing Kit Recalled for Cuvette Failure
FDA (Devices) · 2018-06-13
- HighMedical Device Recall: CARDIOHELP HLS Set Advanced Sterility Barrier Breach
FDA (Devices) · 2022-09-28
- HighHLS Set Advanced Sterility Breach Recall for Cardiac Support Equipment
FDA (Devices) · 2022-09-28