The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1582-2019·Announced 2019-06-05

Maquet Getinge Custom Tubing Sets Recalled Due to Sterility Assurance Concerns

Maquet Cardiovascular is recalling specific custom tubing sets for extracorporeal circulation because a design feature may prevent proper sterilization, potentially compromising sterility.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a potential failure of sterilization assurance (SAL) for a medical device used in extracorporeal circulation, which poses a risk of infection or harm, but no specific injuries or illnesses are reported in the source text.

Plain-English summary

Maquet Cardiovascular, LLC is recalling specific lots of Maquet Getinge-BEQ-T 8068 1/4" WITH BRIDGE custom tubing sets used for extracorporeal circulation. The recall is due to a potential defect in the sterilization process. During sterilization, the closed stopcock and non-vented cap configuration may prevent the flow of humidity and ethylene oxide gas into the fluid path between the closed port and the cap. As a result, the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured.

The affected product is distributed nationwide. The specific batch numbers involved in this recall are 3000041327, 3000042148, 3000042819, 3000046732, 3000054315, 3000055642, 3000062038, and 3000066689. The material code for the product is 701066883.

Healthcare facilities and users should check their inventory for these specific batch numbers. If the product is found, it should be removed from use and returned to the manufacturer or disposed of according to local regulations. This recall is classified as Class II by the FDA, indicating that use of, or exposure to, the violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

The recalled product

Product
Maquet Getinge-BEQ-T 8068 1/4" WITH BRIDGE Material: 701066883

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 183 recalled by Maquet Cardiovascular, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 183 products in this recall →