CareFusion Recalls Alaris Pump 8100 Due to Software Delay Issue
CareFusion is recalling Alaris Pump model 8100 units with specific software versions because a defect may prevent the pump from properly delaying infusions.
The agency’s classification
FDA (Devices): Class I
Class I is a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 4/5 (Severe).
The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.
Our severity reasoning: The agency classified this recall as Class I — a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death. On that classification alone this is scored Severe; a score of Critical is reserved for recalls where a death or serious injury was reported. The agency gives the reason as: "CareFusion is recalling the Alaris Pump model 8100 version 9.1.18 because it may have a software issue that results in situation where the pump module will not properly delay an infusion when the "Delay Until" option or "Multidose" feature…"
Plain-English summary
CareFusion 303, Inc. is recalling the Alaris Pump model 8100 with software version 9.1.18 and associated software upgrade kits. The recall involves 57,219 units in total, with 55,756 units distributed in the United States. The affected products were distributed nationwide in the US as well as in Canada, the United Arab Emirates, Australia, and Kuwait.
The recall is being conducted because the pump may have a software issue that results in the module not properly delaying an infusion when the "Delay Until" option or "Multidose" feature is used. This defect could affect the timing of medication or fluid delivery to patients.
Healthcare facilities and users should check their Alaris Pump model 8100 units against the list of affected serial numbers provided in the official recall notice. If a unit is affected, users should follow the instructions provided by CareFusion to address the software issue.
The recalled product
- Product
- Alaris Pump model 8100 with software version 9.1.18 and software upgrade kits with "Delay Until" Option and "Multidose" Feature The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids…
- Manufacturer
- CareFusion 303, Inc.
- Category
- Medical Device — Infusion Pump
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · CareFusion 303, Inc.
See all →- HighBD Pyxis MedBank Mini fingerprint scanner may overheat and burn
FDA (Devices) · 2026-06-10
- HighBD Pyxis MedStation ES fingerprint scanner may overheat and burn
FDA (Devices) · 2026-06-10
- HighBD Pyxis Anesthesia Station fingerprint scanner may overheat and burn
FDA (Devices) · 2026-06-10
- HighBD Pyxis MedBank Tower fingerprint scanner may overheat and burn
FDA (Devices) · 2026-06-10
- HighBD Pyxis Mini Main cabinets may overheat at fingerprint scanner
FDA (Devices) · 2026-06-10
Same hazard · software defect
See all →- ModerateBeckman Coulter Recalls DxC 500i Clinical Analyzer Modules Due to Software Error
FDA (Devices) · 2026-10-14
- HighPhilips Spectral CT Systems Recalled for Software Issues Affecting Safety
FDA (Devices) · 2026-10-14
- ModerateSteris HexaVue Focus Integration System Recalled for Patient Data Mismatch
FDA (Devices) · 2026-10-14
- ModerateUnited Imaging uMI Panvivo S PET/CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
- ModerateUnited Imaging uCT 550 CT Systems Recalled for Software Defect
FDA (Devices) · 2026-10-07
Similar recalls
- SevereCareFusion Alaris Pump Module Recalled for Software Errors
FDA (Devices) · 2020-03-11
- HighCareFusion Recalls Alaris Pump Modules Due to Flow Rate Variations
FDA (Devices) · 2017-10-25
- HighCareFusion Recalls Alaris Syringe Modules Due to Software Anomaly
FDA (Devices) · 2016-10-05
- SevereCareFusion Recalls Alaris Pump Module 8100 Due to Motor Stalls
FDA (Devices) · 2012-08-22
- HighCareFusion Alaris PC Unit Infusion Pump Software Recall
FDA (Devices) · 2015-06-24
- SevereCareFusion Alaris Syringe Module Software Error Recall
FDA (Devices) · 2020-03-11