The Recall Desk
Agency classClass IIOur severity:HighFDA (Devices)·Z-1549-2019·Announced 2019-06-05

Maquet Getinge ECC Tubing Sets Recalled for Sterility Assurance Concerns

Maquet Cardiovascular is recalling specific batches of Custom Tubing Sets for Extracorporeal Circulation because a design feature may prevent sterilization gas from reaching all parts of the device.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 3/5 (High).

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall involves a medical device where sterility cannot be assured, posing a risk of infection or harm, but the source text does not report any specific illnesses or injuries, fitting the High severity criteria for risk-of-harm products without reported injury.

Plain-English summary

Maquet Cardiovascular, LLC is recalling specific batches of Custom Tubing Sets for Extracorporeal Circulation (ECC) (Material: 701062473). The affected batch numbers are 3000045173, 3000046396, and 3000046746. The products were distributed nationwide.

The recall was initiated because the closed stopcock and non-vented cap configuration may prevent the flow of humidity and ethylene oxide gas into the fluid path between the closed port and the cap during sterilization. As a result, the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured.

Healthcare facilities and users should stop using the affected products and contact Maquet Cardiovascular for further instructions on the recall.

The recalled product

Product
Maquet Getinge-BEQ-T 39700 ECC PACK Material:701062473

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Batch Numbers: 3000045173 3000046396 3000046746

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 183 recalled by Maquet Cardiovascular, LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 183 products in this recall →

Similar recalls