GEM Premier 5000 PAK part 00055407510 may eject during warm-up
Instrumentation Laboratory is recalling GEM Premier 5000 PAK part 00055407510 after confirmed complaints of Process Control Solution Not Detected errors during warm-up causing cartridge ejection.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Rubric level 3 covers risk-of-harm products where injury has not yet been reported: the source states results may be delayed and patient management may require reassessment, and no injuries are stated.
Plain-English summary
Instrumentation Laboratory is recalling the GEM Premier 5000 PAK, part number 00055407510. About 1,126 units were distributed worldwide, covering the United States nationwide and more than sixty other countries. The FDA has classified this recall as Class II.
Confirmed customer complaints indicate that GEM PAKs may experience an increased incidence of "Process Control Solution Not Detected" errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. Consecutive ejections may prolong turnaround times, potentially delaying results, and patient management may require reassessment once results are available.
Affected product carries UDI 08426950807636, across 40 lots beginning 251114J.
Laboratories can identify affected cartridges by the part number, UDI and lot numbers above, and may contact Instrumentation Laboratory for further information.
The recalled product
- Product
- GEM Premier 5000 PAK; Part No. 00055407510.
- Manufacturer
- Instrumentation Laboratory
- Affected units
- 1,126
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Part number: 00055407510
- UDI: 08426950807636
- Lots No.: 251114J
- 251117AC
- 251117T
- 251118B
- 251119B
- 251121P
- 251124O
- 251124Q
- 251126E
- 251201U
- 251201V
- 251201W
- 251202G
- 251203H
- 251203L
- 251204N
- 251204O
- 251205L
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 20 recalled by Instrumentation Laboratory under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 20 products in this recall →Related recalls
Same manufacturer · Instrumentation Laboratory
See all →- HighGEM Premier 7000 with iQM3 Blood Analysis Systems Recalled for Solution Concentration Issues
FDA (Devices) · 2026-08-12
- SevereGEM Premier 5000 PAK Lysing Solution Concentration Recall
FDA (Devices) · 2026-08-12
- HighPortable Blood Analyzer System Recalled Due to Barcode Reader Misidentification Risk
FDA (Devices) · 2026-08-05
- HighGEM Premier 5000 Blood Analyzer Recalled Over Bar Code Reader Patient ID Risk
FDA (Devices) · 2026-08-05
- HighBar Code Reader Wands for GEM Premier Blood Analyzers Recalled Due to Misidentification Risk
FDA (Devices) · 2026-08-05
Same hazard · cartridge ejection
See all →- HighGEM Premier 5000 cartridges may eject repeatedly during analyzer warm-up
FDA (Devices) · 2026-03-18
- HighGEM Premier 5000 PAK part 00055360004 may eject during warm-up
FDA (Devices) · 2026-03-18
- HighGEM Premier 5000 PAK part 00055360011 may eject during warm-up
FDA (Devices) · 2026-03-18
- HighGEM Premier 5000 PAK part 00055360010 may eject during warm-up
FDA (Devices) · 2026-03-18
- HighGEM Premier 5000 PAK part 00055407508 may eject during warm-up
FDA (Devices) · 2026-03-18