Ethicon Recalls Echelon 60mm Endoscopic Linear Cutter Reloads
Ethicon Endo-Surgery is voluntarily recalling Echelon 60mm endoscopic linear cutter reloads worldwide due to potential reload damage that could cause incomplete staple line formation.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk of harm (incomplete staple line formation) in a medical device where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
Ethicon Endo-Surgery Inc is initiating a voluntary worldwide recall of ECHELON 60mm Endoscopic Linear Cutter Reloads (ECR60T). The recall affects 57,540 units of the black reloads packaged in flexible blisters with Tyvek lids. These sterile, single-patient-use instruments are designed to simultaneously cut and staple tissue during endoscopic procedures.
The recall is being conducted due to the potential for incomplete staple line formation resulting from reload damage during the firing sequence. The affected lots have expiration dates ranging from May 15, 2013, to March 15, 2014. The products were distributed nationwide in the United States and in numerous countries including Canada, Belgium, India, Malaysia, Saudi Arabia, Brazil, Israel, Mexico, Singapore, Jordan, New Zealand, Taiwan, Chile, Korea, Puerto Rico, Turkey, Colombia, Kuwait, Qatar, United Arab Emirates, Egypt, Lebanon, and Russia.
Healthcare providers and facilities should stop using the affected reloads and contact Ethicon Endo-Surgery for further instructions. Consumers should not use these medical devices, as they are intended for professional medical use only.
The recalled product
- Product
- ECHELON 60mm Endoscopic Linear Cutter Reloads Black in Flexible Blister pack with Tyvek lid. The ECHELON and ECHELON FLEX families of Endoscopic Linear Cutters (articulating and straight) are sterile, single patient use instruments that simultaneously cut and staple tissue.…
- Manufacturer
- Ethicon Endo-Surgery Inc
- Affected units
- 57,540
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- J4C00J 6/15/2013
- J4C87R 7/15/2013
- J4CC1F 8/15/2013
- J4CD2X 8/15/2013
- J4CF3D 9/15/2013
- J4CJ0E 9/15/2013
- J4CJ78 9/15/2013
- J4CK2P 9/15/2013
- J4CM9X 10/15/2013
- J4CR2H 10/15/2013
- J4CT5H 11/15/2013
- J4CU86 11/15/2013
- J4CV1W 11/15/2013
- J4CV5H 11/15/2013
- K4C149 12/15/2013
- K4C17W 12/15/2013
- K4C33U 12/15/2013
- K4C41C 12/15/2013
- K4C600 1/15/2014
- K4C64A 1/15/2014
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Ethicon Endo-Surgery Inc
See all →- HighLinear cutter reloads recalled after devices were shipped unsterilized
FDA (Devices) · 2025-10-22
- CriticalEndopath Echelon Vascular White Reload surgical stapler device lockout
FDA (Devices) · 2025-05-28
- HighEndoscopic Surgical Stapler Reload May Fail to Form Complete Staple Line
FDA (Devices) · 2024-02-07
- HighHARMONIC HD 1000i Surgical Shears Recalled for Potential Burn Risk
FDA (Devices) · 2021-09-01
- HighSurgical Shears HARMONICHD 1000i Recalled Due to Component Defect and Burn Risk
FDA (Devices) · 2021-09-01
Same hazard · incomplete staple line
See all →- HighTri-Staple 2.0 surgical stapler cartridge damage may cause incomplete stapling
FDA (Devices) · 2024-06-05
- HighIntuitive Surgical Recalls daVinci XI Sureform 60 Reloads Due to Staple Deployment Issues
FDA (Devices) · 2020-05-13
- HighIntuitive Surgical Recalls daVinci XI Sureform Stapler Reloads Due to Deployment Defect
FDA (Devices) · 2020-05-13
- SevereEthicon Endo-Surgery Recalls Echelon Vascular White Reloads Due to Staple Line Defect
FDA (Devices) · 2015-05-27
Similar recalls
- SevereEthicon Endo-Surgery Recalls Echelon Vascular White Reloads Due to Staple Line Defect
FDA (Devices) · 2015-05-27
- HighLinear cutter reloads recalled after devices were shipped unsterilized
FDA (Devices) · 2025-10-22
- HighEndoscopic Surgical Stapler Reload May Fail to Form Complete Staple Line
FDA (Devices) · 2024-02-07
- SevereEthicon Recalls ECHELON FLEX Staplers Due to Anvil Defect
FDA (Devices) · 2019-11-06
- SevereEthicon Echelon Flex Endopath Stapler Recall for Out-of-Spec Anvil
FDA (Devices) · 2019-11-06
- HighEthicon Endopath ETS-FLEX Endoscopic Linear Cutter Recalled for Gear Failure Risk
FDA (Devices) · 2017-06-07