Siemens Multix Fusion Radiographic System Recall for Loose Hose Holder
Siemens is recalling 12 Multix Fusion radiographic systems because a corrugated hose holder may become loose. The issue requires verification and securing of the holder.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: This is a Class II recall involving a mechanical component that may become loose, but no injuries or illnesses are reported, fitting the criteria for a moderate severity voluntary precautionary recall.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 12 units of the Multix Fusion radiographic system. This system is used in hospitals, clinics, and medical practices for whole-body radiographic exposures. The recall involves model numbers 10746665, 10746666, 10847610, and 10746700.
The recall was initiated because the corrugated hose holder located above the 3D rail on the system was observed to become loose. According to the installation instructions, this holder should be affixed with a transversal carriage, Loctite 243, and a torque of 10nm during installation. Siemens has initiated an update instruction to verify and secure the hose holder with Loctite 243.
The affected units were distributed in the states of Alaska, Pennsylvania, Kentucky, Ohio, Wisconsin, Michigan, Florida, and New York. Healthcare facilities using these systems should contact Siemens to verify the status of the hose holder and ensure it is properly secured.
The recalled product
- Product
- MULTIX FUSION. The Multix Fusion system is a radiographic system used in hospitals, clinics, and medical practices. Multix Fusion enables radiographic exposures of the whole body including: skull, chest, abdomen, and extremities and may be used on pediatric, adult and bariatri
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Hazard
- Affected units
- 12
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (4)
- model nos. 10746665
- 10746666
- 10847610
- 10746700
Distribution
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · loose component
See all →- HighRitchey Recalls Carbon Bicycle Forks Due to Fall and Crash Hazard
CPSC · 2026-08-13
- SevereGiraffe infant resuscitation system may deliver wrong oxygen concentration
FDA (Devices) · 2026-06-24
- SevereGiraffe and Panda warmer blender knobs may deliver wrong oxygen concentration
FDA (Devices) · 2026-06-24
- SeverePanda freestanding warmer blender knobs may deliver wrong oxygen concentration
FDA (Devices) · 2026-06-24
- HighPhoroptor VRx Digital Refraction System head may become loose or detach
FDA (Devices) · 2025-04-09