Zimmer Biomet Alvarado Knee Holder Recalled for Foreign Material
Zimmer Surgical Inc is recalling Alvarado and Alvarado II knee holder components due to potentially inadequate cleaning procedures that may leave foreign material on the devices.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a risk of harm due to foreign material, but the source text does not report any specific injuries or illnesses, fitting the criteria for a high-risk product where injury has not yet been reported.
Plain-English summary
Zimmer Surgical Inc is voluntarily recalling the Alvarado Knee Holder Base Plate Assembly and Foot Piece, as well as the Alvarado II Base Plate and Foot Piece. The recall is being conducted because potentially inadequate cleaning procedures may have resulted in foreign material remaining on the devices.
The affected products include the Alvarado Foot Piece (Item Number 00-1320-010-00), Alvarado Knee Holder Base Plate Assembly (Item Number 00-1320-011-00), Alvarado II Foot Piece (Item Number 00-1320-210-00), and Alvarado II Base Plate (Item Number 00-1320-211-00). A total of 37,402 units are involved in this recall, which covers all lot codes. These devices are designed to hold a patient's knee in a selected position during surgery.
The products were distributed worldwide, including in the United States and various international locations such as Australia, Brazil, Canada, and Japan. Healthcare facilities and distributors should stop using the affected devices and contact Zimmer Surgical Inc for further instructions on the recall process.
The recalled product
- Product
- Zimmer Biomet Alvarado Knee Holder Base Plate Assembly and Foot Piece, & Alvarado II Base Plate and Foot Piece: Alvarado Foot Piece: Item Number: 00-1320-010-00 Alvarado Knee Holder Base Plate Assembly: Item Number: 00-1320-011-00 Alvarado II Foot Piece: Item Number: 00-1320-
- Manufacturer
- Zimmer Surgical Inc
- Affected units
- 37,402
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- All lot codes
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · Zimmer Surgical Inc
See all →- ModerateSterile Tourniquet Hoses Recalled Due to Potential Sterility Breach
FDA (Devices) · 2026-07-08
- ModerateSterile Tourniquet Hoses recalled for potential sterility breach
FDA (Devices) · 2026-07-08
- ModerateTourniquet systems recalled over screens that may freeze in non-English settings
FDA (Devices) · 2026-02-04
- ModerateA.T.S. 3200TS tourniquets recalled over screens that may freeze
FDA (Devices) · 2026-02-04
- ModerateDermatome AN units recalled because the thickness control bar may be misaligned
FDA (Devices) · 2025-12-24
Same hazard · foreign material
See all →- ModerateGolden State Medical Supply Recalls Carbamazepine Tablets Due to CGMP Deviations
FDA (Drugs) · 2026-08-26
- HighSun Pharmaceutical Recalls Carbamazepine Tablets Due to Manufacturing Deviations
FDA (Drugs) · 2026-08-26
- HighDQ Chocolate Reduced Fat Ice Cream Mix Recalled Due to Metal Shavings
FDA (Food) · 2026-08-05
- SevereBaked spaghetti and chicken cobbler recalled for possible metal fragment contamination
USDA FSIS · 2026-07-31
- HighRuby Artichoke Hearts Quartered Recalled for Glass Contamination
FDA (Food) · 2026-07-22