Wright Medical Recalls Crosscheck Plating System Due to Dimensional Defects
Wright Medical Technology is recalling 52 units of the Crosscheck Plating System because dimensional specifications are out of range, posing a risk of intraoperative malfunction.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a risk of intraoperative malfunction, which is a risk-of-harm product where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
Wright Medical Technology, Inc. is recalling 52 units of the CROSSCHECK(R) Plating system, specifically the HOCKEY STICK CX Plate, Right (REF CCP-HSX1R), which is used for orthopedic purposes. The affected units were distributed nationwide.
The recall was initiated because the minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification. These dimensional deviations may result in intraoperative malfunction during surgical procedures.
Healthcare providers and facilities that have received these specific lot numbers (10346, 10346-1, 35049-23A-1) should stop using the affected devices and contact Wright Medical Technology for instructions on return or replacement.
The recalled product
- Product
- CROSSCHECK(R) Plating system, HOCKEY STICK CX Plate, Right, REF CCP-HSX1R. Orthopedic use.
- Manufacturer
- Wright Medical Technology, Inc.
- Affected units
- 52
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (3)
- Lot Numbers: 10346
- 10346-1
- 35049-23A-1
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 19 recalled by Wright Medical Technology, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 19 products in this recall →Related recalls
Same manufacturer · Wright Medical Technology, Inc.
See all →- HighWright Medical Recalls Charlotte Claw Screws Due to Missing Threads
FDA (Devices) · 2026-10-07
- HighStryker Infinity Resection Adjustment Block Recall: Missing Locking Screw
FDA (Devices) · 2024-04-10
- HighStryker Infinity Alignment Frame Distal Sub-Assembly Recalled for Manufacturing Defect
FDA (Devices) · 2023-07-12
- HighStryker ORTHOLOC 3Di Fusion Plate Recalled for Incorrect Marked Descriptions
FDA (Devices) · 2023-05-03
- HighStryker ORTHOLOC 3Di Fusion Plate Contains Incorrect Laser Markings
FDA (Devices) · 2023-05-03
Same hazard · intraoperative malfunction
See all →- HighWright Medical Recalls CrossCHECK Non-Locking Screws Due to Malfunction Risk
FDA (Devices) · 2015-09-09
- HighWright Medical CrossCHECK Locking Screw Recall for Intraoperative Malfunction Risk
FDA (Devices) · 2015-09-09
- HighWright Medical Recalls Crosscheck Plating System Due to Specification Defects
FDA (Devices) · 2015-04-29
- HighWright Medical Recalls Crosscheck Plating System Due to Dimensional Defects
FDA (Devices) · 2015-04-29
- HighWright Medical Recalls Crosscheck Plating System Due to Spec Failures
FDA (Devices) · 2015-04-29
Similar recalls
- HighWright Medical Recalls CrossCHECK Non-Locking Screws Due to Malfunction Risk
FDA (Devices) · 2015-09-09
- HighWright Medical CrossCHECK Locking Screw Recall for Intraoperative Malfunction Risk
FDA (Devices) · 2015-09-09
- HighWright Medical Recalls Orthopedic Claw Plates Due to Incorrect Raw Material
FDA (Devices) · 2014-09-17
- ModerateWright Medical Recalls CLAW II Orthopedic Plates Due to Material Error
FDA (Devices) · 2014-09-17
- HighWright Medical Recalls CLAW II Straight Plate OrthoLoc 3DSi Orthopedic Plates
FDA (Devices) · 2014-09-17
- HighWright Medical Recalls 3.5mm Claw Plates Due to Incorrect Raw Material
FDA (Devices) · 2014-09-17