TZ Medical Recalls Defibrillation Electrodes Due to Defective White Wire
TZ Medical Inc. is recalling P-211-M1 and P-214-M1 defibrillation electrodes because they may contain a defective white wire on the connector assembly.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device used for life-critical procedures (defibrillation/pacing) with a potential functional defect, but the source text does not report any injuries, illnesses, or fatalities, fitting the 'High' criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
TZ Medical Inc. is recalling its P-211-M1 and P-214-M1 Medtronic / Physio Control Compatible transparent defibrillation electrodes. The recall is being conducted because the products may contain a defective white wire on the connector assembly.
These disposable electrodes are used by trained professionals in hospitals, doctors' offices, and Emergency Medical Services for adult low-energy defibrillation, transcutaneous pacing, cardioversion, and monitoring. The affected units were distributed nationwide in the United States and in Chile.
If a failure occurs, the end user can identify the issue by inspecting the white wire located 1/4 inch from the Gray Medtronic / Physio adapter. Consignees will notice discoloration on the wire. Users should inspect their equipment for this specific defect.
The recalled product
- Product
- adult defibrillation, cardioversion, pacing, monitoring Disposable Electrodes, Model P-211-M1 and P-214-M1. Product Usage: Used by trained professionals in hospitals, doctors offices, and Emergency Medical Services for adult low-energy defibrillation, transcutaneous pacing, c
- Manufacturer
- TZ Medical Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (14)
- Model P-211-M1 Lot numbers: Y081712-17
- Y081712-18
- Y081712-19
- Y081712-20
- Y112612-08
- Y112612-09
- Y112612-10
- Y112612-11
- Y112612-12
- Y112612-13
- and Y020113-07. Model P-214-M1 Lot numbers: Y081712-04
- Y112612-24
- Y112612-23
- and Y031913-10.
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · TZ Medical Inc.
See all →- ModerateTZ Medical Recalls Mis-Labeled Neptune Plus Hemostatic Pads
FDA (Devices) · 2016-11-02
- ModerateTZ Medical Recalls SilverStat Hemostasis Pads Due to Labeling Discrepancy
FDA (Devices) · 2015-11-25
- ModerateTZ Medical Neptune Plus Hemostasis Pad Recall for Labeling Mismatch
FDA (Devices) · 2015-11-25
Same hazard · electrical failure
See all →- HighHalcore Recalls 2025-2026 Horton Ambulances Due to Touchscreen Failure
NHTSA · 2026-08-14
- HighMarion Recalls 2025 Pumper Fire Engines Due to Display Unit Failure
NHTSA · 2026-08-10
- HighRosenbauer Recalls Emergency Vehicles with Faulty Display Units
NHTSA · 2026-08-03
- SevereBMW S 1000 RR Motorcycles Recalled Due to Potential Engine Stall Risk
NHTSA · 2026-07-23
- High