Elekta Integrity 3.0 Multileaf Collimator Recalled for Beam Delivery Tolerance Issues
Elekta, Inc. is recalling Integrity 3.0 multileaf collimators because incorrect static tolerances may prevent the detection of unexpected movement during radiation beam delivery.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device where a failure to detect unexpected movement during radiation delivery poses a risk of harm, but the source text does not report any specific injuries or illnesses, fitting the criteria for a risk-of-harm product where injury has not yet been reported.
Plain-English summary
Elekta, Inc. is recalling 31 units of the Integrity 3.0 multileaf collimator. This device is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed. The recall is due to a software issue where incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0 may prevent the detection of unexpected movement during beam delivery.
The affected units were distributed nationwide in the United States, including California, Florida, Illinois, Maryland, Massachusetts, New Jersey, New York, North Carolina, North Dakota, Ohio, Oregon, Pennsylvania, Tennessee, Texas, Virginia, Washington, the District of Columbia, and Puerto Rico. Specific serial numbers for the affected units are listed in the agency's recall record.
Healthcare providers and facilities using these devices should contact Elekta, Inc. for instructions on how to address the issue. The source text does not report any specific injuries or illnesses associated with this recall.
The recalled product
- Product
- Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.
- Manufacturer
- Elekta, Inc.
- Affected units
- 31
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- 325134
- 325254
- 325082
- 325159
- 325268
- 325237
- 325331
- 325267
- 325236
- 325314
- 325127
- 325261
- 325286
- 325077
- 325269
- 325129
- 325296
- 325273
- 325139
- 325201
Distribution
Related recalls
Same manufacturer · Elekta, Inc.
See all →- SevereFletcher CT/MR Shielded Applicator Set for Gynecological Brachytherapy Recalled
FDA (Devices) · 2026-07-15
- SevereBrachytherapy Applicator Advanced Gynecological Applicator Tip Detachment
FDA (Devices) · 2026-07-15
- SevereElekta Geneva Brachytherapy Applicator Intrauterine Tube Detachment Risk
FDA (Devices) · 2026-07-15
- HighMOSAIQ oncology software may fail to load adapted treatment plan
FDA (Devices) · 2026-06-17
- HighLeksell Vantage Arc System locking mechanism may not secure properly
FDA (Devices) · 2026-05-06
Same hazard · radiation exposure
See all →- HighPhilips Ingenuity CT Scanners Recalled Due to Bolus Tracking Trigger Issue
FDA (Devices) · 2026-08-05
- HighPhilips CT Scanners Recalled Due to Bolus Tracking Trigger Issue
FDA (Devices) · 2026-08-05
- HighPhilips IQon Spectral CT Recalled Due to Automatic Scan Trigger Failure
FDA (Devices) · 2026-08-05
- HighPhilips Allura Xper hand switch buttons may not fully release
FDA (Devices) · 2026-06-24
- HighPhilips MultiDiagnost Eleva hand switch buttons may not fully release
FDA (Devices) · 2026-06-24