Synthes Recalls SYNTHECEL Dura Repair Due to Adhesion Risks
Synthes (USA) Products LLC is recalling 200 units of SYNTHECEL Dura Repair because the device may adhere to anatomical structures, posing risks if removal is attempted.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a high-risk medical device where a potential risk of harm (adhesion leading to injury during removal) is identified, but no specific injuries or illnesses have been reported in the source text.
Plain-English summary
Synthes (USA) Products LLC is recalling 200 units of SYNTHECEL Dura Repair, a Class II medical device intended for use as a dura replacement in adults. The recall involves specific lot numbers and part numbers for the 2.5 cm x 7.5 cm size. The product was distributed nationwide to accounts in 32 states, with no foreign distribution reported.
The recall was initiated because it was reported that the SYNTHECEL Dura Repair could adhere to various anatomical structures after implantation. This adhesion may expose patients to potential risks if a surgeon intends to temporarily apply and subsequently remove the implant. The device is intended to remain in place and is not designed for removal.
To address this issue, Synthes is updating the labeling and promotional material for SYNTHECEL Dura Repair to clarify that the device is intended to remain in place and is not intended to be removed. Healthcare providers and patients should be aware of this updated indication regarding the permanence of the implant.
The recalled product
- Product
- SYNTHECEL Dura Repair, 510K # K131792, K113071, Class II, Classification name: Dura Substitute Device listing number: D140596, D211012 SYNTHECEL Dura Repair is intended for use as a dura replacement for the repair of dura mater in adults.
- Manufacturer
- Synthes (USA) Products LLC
- Category
- Medical Device — Dura Substitute
- Affected units
- 200
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (9)
- SYNTHECEL Dura Repair 2.5 cm x 7.5 cm (1x3)
- part numbers: SC.400.019.01S
- Lot Numbers: 7840492
- 7855337
- 7860565
- 7880949
- 7901437
- 7909570
- 7922341
Distribution
Part of a larger recall action
This product is one of 5 recalled by Synthes (USA) Products LLC under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 5 products in this recall →Related recalls
Same manufacturer · Synthes (USA) Products LLC
See all →- HighVOLT locking screw taps swapped between lots carry incorrect thread
FDA (Devices) · 2026-04-08
- HighVOLT bone taps swapped between lots carry the incorrect thread
FDA (Devices) · 2026-04-08
- ModerateLocking bone screws recalled after packaging labeled with the wrong length
FDA (Devices) · 2025-10-29
- ModerateForty six millimetre locking screws recalled over wrong length on packaging
FDA (Devices) · 2025-10-29
- HighSurgical screwdrivers recalled for failure to engage with implant screws
FDA (Devices) · 2024-06-05
Same hazard · adhesion
See all →- HighHemospray Endoscopic Hemostat Recall Due to Endoscope Adhesion Risk
FDA (Devices) · 2023-06-28
- HighZimmer Biomet Recalls Periarticular Shoulder Plates Due to Packaging Adhesion
FDA (Devices) · 2019-11-13
- ModerateZimmer Biomet Recalls Various Knee Systems Due to Packaging Adhesion
FDA (Devices) · 2019-11-13
- ModerateZimmer Biomet Recalls Fitmore Hip Stems Due to Packaging Adhesion
FDA (Devices) · 2019-11-13
- ModerateZimmer Biomet Recalls Universal Locking Plates Due to Packaging Adhesion
FDA (Devices) · 2019-11-13
Similar recalls
- HighSynthes Radial Head Prosthesis Recalled for Potential Stem Loosening
FDA (Devices) · 2017-02-08
- HighSynthes Flexible Medullary Reamer Recalled Due to Cleaning and Corrosion Risks
FDA (Devices) · 2013-08-21
- HighSynthes Tensioning Device Recalled Due to Potential Mechanical Failure
FDA (Devices) · 2013-12-04
- ModerateSynthes Recalls 2.0mm Stardrive Screwdriver for Incorrect MR Safety Labeling
FDA (Devices) · 2016-07-20
- ModerateSynthes Recalls 34 Stardrive Screwdrivers for MR Safety Labeling Errors
FDA (Devices) · 2016-07-20
- HighSynthes RIA Tube Assembly Recalled for Unconfirmed Sterility
FDA (Devices) · 2024-03-13