Agency classClass IIOur severity:SevereFDA (Devices)·Z-1370-2015·Announced 2015-04-08

BD FACSCalibur Instruments Recalled for Incorrect Circuit Breaker Amperage

BD Biosciences is recalling 72 BD FACSCalibur instruments because some have circuit breakers with incorrect amperage ratings, which may fail to open during over-current situations.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall is classified as FDA Class II, and the defect involves a circuit breaker failure that could lead to electrical hazards or fire, meeting the criteria for a Severe rating.

Plain-English summary

BD Biosciences, Systems & Reagents is recalling 72 units of the BD FACSCalibur instrument, used for in vitro diagnostics. The recall involves specific catalog numbers (342973, 342975, 34976) and is identified by recall number Z-1370-2015.

The recall is due to a defect where the circuit breakers on some of these instruments may have an incorrect rated amperage of 20 amps instead of the required 2 amps. This discrepancy may cause the circuit breaker to fail to open in an over-current situation.

The affected devices were distributed worldwide, including in the United States (Texas, New Jersey, and Oklahoma) and internationally to Hong Kong, Japan, Singapore, and Belgium. Users should check their device serial numbers against the list provided in the official recall notice to determine if their unit is affected.

The recalled product

Product
BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.
Affected units
72

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • Serial Nos. E34297300656
  • E34297300657
  • E34297300659
  • E34297300660
  • E34297300661
  • E34297300666
  • E34297300667
  • E34297300668
  • E34297300669
  • E34297300670
  • E34297300671
  • E34297300673
  • E34297300675
  • E34297300676
  • E34297300677
  • E34297300678 E34297502563
  • E34297502564
  • E34297502565
  • E34297502566
  • E34297502567

Distribution

Distributed in 3 states: