Dental bonding primer may lack resin component affecting restoration
Ultradent Peak Universal Bond self-etch dental primers may be missing resin, potentially compromising bond strength. Affected products were distributed nationwide and internationally.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: FDA Class II medical device with no reported hospitalizations or injuries. The recall addresses a risk-of-harm product (dental bonding integrity) where injury has not yet been reported, meeting the criterion for Score 3 per the severity rubric.
Plain-English summary
Ultradent Products, Inc. is recalling specific lots of Peak Universal Bond Self-Etch dental bonding products, including the Bottle Kit (Part 4541), Intro Kit (Part 4554), SE Primer Refill (Part 5135), and Sample Kit (Part S4554). These products are used by dentists for light-curing dental bonding procedures. The recall affects 1,331 kits comprising 5,009 syringes.
The recalled products may be missing the resin component of the self-etch primer chemistry. This defect could compromise bond strength and affect the integrity of dental restorations.
The products were distributed worldwide. In the United States, distribution included Florida, Michigan, Nevada, Washington, Colorado, Virginia, Puerto Rico, Wisconsin, Iowa, Tennessee, Arkansas, Minnesota, Wyoming, and Maine. International distribution included Kuwait, Italy, France, Canada, Spain, Malaysia, Russia, and Greece.
Dentists who have received affected products should verify lot numbers against the recall list. Products matching the affected lot numbers (BLRCR, BM1Y7, BM9G3 for Part 4541; BLSGH, BM578, BN449 for Part 4554; BLX3F, BLSGM, BLXSG, BLXW4, BM34M, BM784, BM9KB, BMJVK, BMTZM for Part 5135; and BLRJJ, BM1NH for Part S4554) should not be used. Contact Ultradent Products, Inc. for replacement or further instructions.
The recalled product
- Product
- Peak Universal Bond Self-Etch Bottle Kit, Part: 4541; Peak Universal Bond Self-Etch Intro Kit, Part: 4554; Peak SE Primer 4pk Refill, Part: 5135; Peak Universal Bond Sample SE, Part: S4554. For use in light-curing dental bonding procedures.
- Manufacturer
- Ultradent Products, Inc.
- Category
- Medical Device — Dental Bonding
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (17)
- REF/ UDI-DI/ Lot: 4541/ 00883205104274/ BLRCR
- BM1Y7
- BM9G3
- 4554/ 00883205104281/ BLSGH
- BM578
- BN449
- 5135/ 00883205104311/ BLX3F
- BLSGM
- BLXSG
- BLXW4
- BM34M
- BM784
- BM9KB
- BMJVK
- BMTZM
- S4554/ 00883205102065/ BLRJJ
- BM1NH
Distribution
Related recalls
Same manufacturer · Ultradent Products, Inc.
See all →- HighJiffy composite polisher cups may crumble and break apart easily
FDA (Devices) · 2026-04-08
- ModerateUltradent Recalls Vit-l-escence Composite Due to Red Particulates
FDA (Devices) · 2024-09-18
- HighDental Composite Resin Recall Due to Red Particulate Contamination
FDA (Devices) · 2024-09-18
- HighFDA Recalls Transcend Dental Composite Material Due to Manufacturing Defect
FDA (Devices) · 2024-09-18
- ModerateUltradent PermaFlo Flowable Composite Recalled for Manufacturing Cross-Contamination
FDA (Devices) · 2020-10-14
Same hazard · improper bonding
See all →- HighCAPIOX FX25 Oxygenator Recalled for Potential Leaks from Improper Bonding
FDA (Devices) · 2023-04-12
- HighTerumo Recalls CAPIOX NX19 Oxygenator Due to Improper Bonding
FDA (Devices) · 2023-04-12
- HighCAPIOX RX15 Oxygenator units recalled due to improper thermistor bonding
FDA (Devices) · 2023-04-12
- HighTerumo Recalls Winged Infusion Sets Due to Improper Bonding
FDA (Devices) · 2015-12-30
Similar recalls
- ModerateUltradent Recalls Peak SE Primer Kits Due to Mislabeled Syringes
FDA (Devices) · 2012-06-27
- HighFDA Recalls Transcend Dental Composite Material Due to Manufacturing Defect
FDA (Devices) · 2024-09-18
- LowSeptodont Recalls Compolite Bond SE Due to Incorrect Packaging
FDA (Devices) · 2014-08-06
- High3M Unitek Orthodontic Primer Recalled Due to Bond Failures and Skin Irritation
FDA (Devices) · 2025-01-08
- HighDental Composite Resin Recall Due to Red Particulate Contamination
FDA (Devices) · 2024-09-18
- LowSeptodont Recalls Self-Etch Bond Adhesive Due to Packaging Error
FDA (Devices) · 2014-08-06