Outset Tablo Hemodialysis System Recall Due to Software Manufacturing Defect
Outset Medical is recalling 374 Tablo hemodialysis systems due to a software malfunction introduced during manufacturing. The defect causes frequent high conductivity dialysate alarms in devices distributed nationwide.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: This is an FDA Class II device recall with no reported injuries or illnesses. However, it involves a life-critical hemodialysis system with a software malfunction affecting system alarms, fitting the criterion of a risk-of-harm product where injury has not yet been reported.
Plain-English summary
Outset Medical is recalling 374 Tablo hemodialysis systems (Model Numbers PN-0006000 and PN-0007001). The systems are being recalled due to a software defect introduced during manufacturing that causes an observed trend of frequent high conductivity dialysate alarms.
The Tablo is a self-contained hemodialysis system intended for acute, home, and chronic dialysis therapy, with or without ultrafiltration, in acute or chronic care facilities.
The 374 affected systems were distributed nationwide throughout the United States, including in Alaska, Alabama, Arkansas, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Michigan, Mississippi, Missouri, Montana, Nebraska, Nevada, New Jersey, New Mexico, New York, North Carolina, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Tennessee, Texas, Utah, Vermont, Virginia, and Wisconsin. Specific serial numbers identifying the recalled units are available from Outset Medical and the FDA.
The recalled product
- Product
- Outset Tablo Model Numbers (REF): PN-0006000 PN-0007001 Tablo is a next generation, self-contained hemodialysis system (Hemodialysis System or System), intended for acute, home and chronic dialysis therapy with or without ultrafiltration, in an acute or chronic care…
- Manufacturer
- Outset Medical, Inc.
- Affected units
- 374
Distribution
Related recalls
Same manufacturer · Outset Medical, Inc.
See all →- HighHemodialysis console blood leak sensor may fail to trigger alarm
FDA (Devices) · 2024-10-09
- HighTabloCart power cord connector may overheat in hemodialysis system
FDA (Devices) · 2024-06-26
- SevereTablo Hemodialysis System recalled for PCB contamination in tubing
FDA (Devices) · 2024-04-24
- HighOutset Tablo Hemodialysis Console Recalled for Heat Damage Risk
FDA (Devices) · 2022-03-30
Same hazard · software malfunction
See all →- HighPhilips PIC iX Essentials Patient Information System Software Recall
FDA (Devices) · 2026-09-23
- HighPhilips Patient Information Center iX Software Malfunction Recall
FDA (Devices) · 2026-09-23
- HighAbbott Merlin pacemaker programmer may fail to terminate threshold test
FDA (Devices) · 2026-06-17
- HighMedtronic SynchroMed Flex Infusion Programmer Software Application
FDA (Devices) · 2026-05-20
- ModerateProtocol Configuration Tool software v1.3.0 for Hamilton AutoLys STAR
FDA (Devices) · 2026-04-29
Similar recalls
- HighOutset Tablo Hemodialysis Console Recalled for Heat Damage Risk
FDA (Devices) · 2022-03-30
- SevereTablo Hemodialysis System recalled for PCB contamination in tubing
FDA (Devices) · 2024-04-24
- HighHemodialysis console blood leak sensor may fail to trigger alarm
FDA (Devices) · 2024-10-09
- HighTabloCart power cord connector may overheat in hemodialysis system
FDA (Devices) · 2024-06-26
- HighFresenius K2 Hemodialysis Machines Recalled for SLED Program Software Defect
FDA (Devices) · 2018-04-11
- HighBaxter HomeChoice Pro APD System Recalled for Occluder Defect
FDA (Devices) · 2015-03-25