Biocare Medical IntelliPath FLX Automated Staining Instrument Recall
Biocare Medical is recalling 72 IntelliPath FLX instruments due to a software defect that may cause random movement and buffer dispensing errors.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk of physical damage to the instrument and components, which fits the criteria for a high-risk product where injury has not yet been reported.
Plain-English summary
Biocare Medical, LLC is recalling 72 units of the IntelliPath FLX Automated Staining Instrument. The affected units are Model Numbers IPS0001US and IPS0001INTL containing software version 3.5.3.1. The recall covers units distributed in the United States and internationally to Germany and the Netherlands.
The recall is due to a potential software defect that may cause the instrument to move at random speeds and in random directions across its range of motion. This malfunction could result in the instrument hitting end stops or colliding with other components on the working deck. Additionally, the defect may cause wash buffer to be dispensed in random locations.
Affected units are identified by specific serial numbers listed in the agency's structured data. Consumers and healthcare facilities should contact Biocare Medical, LLC for instructions on how to address the software issue.
The recalled product
- Product
- BIOCARE Medical , IntelliPath FLX, Automated Staining Instrument, Ref: " Model Number:IPS0001US and IPS0001INTL containing software version 3.5.3.1
- Manufacturer
- Biocare Medical, LLC
- Affected units
- 72
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Affected US Serial Numbers: BCIP01-0021
- BCIP01-0033
- BCIP01-0090
- BCIP01-0106
- BCIP01-0112
- BCIP01-0148
- BCIP01-0151
- BCIP01-0187
- BCIP01-0189
- BCIP01-0198
- BCIP01-0201
- BCIP01-0216
- BCIP01-0221
- BCIP01-0249
- BCIP01-0262
- BCIP01-0265
- BCIP01-0266
- BCIP01-0268
- BCIP01-0279
- BCIP01-0285
Distribution
Related recalls
Same manufacturer · Biocare Medical, LLC
See all →- ModerateBiocare Medical Recalls IHC Kits Due to Potential Staining Failure
FDA (Devices) · 2024-11-13
- HighBiocare Medical Recalls 4plus Streptavidin HRP Label Due to Staining Failure
FDA (Devices) · 2020-06-17
- ModerateBiocare Medical Recalls URO-3 Triple Stain Kits Due to Component Mislabeling
FDA (Devices) · 2018-09-26
- ModerateBiocare Medical Recalls Vulcan Fast Red Chromogen Kit Due to Stability Issue
FDA (Devices) · 2018-07-11
- ModerateBiocare Medical Recalls Mislabeled Mach 4 Mouse Probe Vials
FDA (Devices) · 2016-04-27
Same hazard · uncontrolled movement
See all →- HighRebstock Holding Rod for Fixation May Move Uncontrollably
FDA (Devices) · 2024-03-06
- HighSkytron GS70H Surgical Table Recall: Control Failure Causes Continued Movement
FDA (Devices) · 2023-09-06
- HighX-Ray Machine Detector May Move Unexpectedly Due to Drive Shaft Failure
FDA (Devices) · 2023-06-07
- HighMedical Device Recall: X-Ray Detector Drive Shaft May Break
FDA (Devices) · 2023-06-07
- Highda Vinci Surgical System Patient Cart Ballscrew Failure Risk
FDA (Devices) · 2023-03-22
Similar recalls
- HighPhilips Intellivue iX Software Recall for Patient Discharge Failure
FDA (Devices) · 2018-03-07
- ModerateBiocare Medical Recalls IHC Kits Due to Potential Staining Failure
FDA (Devices) · 2024-11-13
- HighPhilips IntelliSpace Portal Software Recall for SPECT Reconstruction Errors
FDA (Devices) · 2014-06-11
- ModeratePhilips Brilliance iCT Recalled for Software Performance Issues
FDA (Devices) · 2016-08-17
- ModeratePhilips DigitalDiagnost Software Recall for X-ray System Stitching Defects
FDA (Devices) · 2018-10-31
- HighMedical imaging system software errors may delay treatment procedures
FDA (Devices) · 2022-04-27