Siemens ACUSON P300 Ultrasound Systems Recalled for Transducer Overheating
Siemens is recalling ACUSON P300 ultrasound systems with PA230 transducers because they may exceed maximum allowable surface temperature limits.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II, and the hazard involves a risk of thermal injury (burn) to patients, which fits the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 675 ACUSON P300 ultrasound systems equipped with the P300 PA 230E cardiac probe. The recall is classified as Class II by the U.S. Food and Drug Administration. The affected devices were distributed worldwide, including in the United States and numerous international markets.
The recall was initiated because the ultrasound systems, when using the PA230 transducer with a user-defined preset that includes the 100% setting for output power, could cause the transducer surface temperature to exceed the maximum allowable limit. This condition presents a risk of thermal injury to patients during use.
Healthcare facilities and users should identify affected units by checking the model number 10433816 for the ACUSON P3000 Ultrasound System and model number 10852384 for the P300 PA 320E transducer, along with the specific serial numbers listed in the official recall notice. Users should contact Siemens Medical Solutions for instructions on corrective actions.
The recalled product
- Product
- Siemens Medical Solutions, ACUSON P300 Ultrasound system with P300 PA 230E cardiac probe (Esaote product 7348). Provides imaging for guidance of biopsy and imaging to assist in the placement of needles in vascular or other anatomical structures as well as peripheral nerve
- Manufacturer
- Siemens Medical Solutions USA, Inc.
- Affected units
- 675
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Model number 10433816 - ACUSON P3000 Ultrasound System
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc.
See all →- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · thermal injury
See all →- HighCASE cardiac testing systems may spark at the printer head
FDA (Devices) · 2026-06-10
- HighEndoscopy video processors recalled over hot light guide and thermal injury risk
FDA (Devices) · 2025-08-27
- HighVideo Colonoscope May Cause Thermal Burns to Tissue During Use
FDA (Devices) · 2025-03-05
- HighStryker iBur Surgical Bur Recall Due to Excessive Heat Risk
FDA (Devices) · 2024-09-04
- HighStryker iBur 3.0mm surgical burr recalled over thermal hazard
FDA (Devices) · 2024-09-04