RayStation Radiation Therapy Software Recalled for Dose Calculation Errors
RaySearch Laboratories is recalling RayStation Radiation Therapy Treatment Planning System software due to an issue with photon dose calculation for Dynamic MLC plans.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device software defect that could lead to incorrect radiation dose calculations, posing a risk of harm to patients, although no specific injuries or illnesses are reported in the source text.
Plain-English summary
RaySearch Laboratories AB is recalling the RayStation Radiation Therapy Treatment Planning System software, specifically versions 3.5, 4.0, 4.5, and 4.7. The recall involves 245 domestic units distributed nationwide. The affected software is used for treatment planning and analysis of radiation therapy.
The recall is due to an issue with photon dose calculation for Dynamic MLC (DMLC) plans on machines where the MLC is positioned above the jaws, such as some Elekta linacs. The magnitude of the error depends on the beam model output factor corrections and the individual DMLC plan characteristics. For RayStation 4.7, the issue applies to machines with fixed jaws regardless of MLC/jaw position.
Healthcare providers using this software should contact RaySearch Laboratories for further instructions on how to address the dose calculation issue.
The recalled product
- Product
- RayStation Radiation Therapy Treatment Planning System; -- RayStation 3.5, RayStation 4.0, RayStation 4.5 and RayStation 4.7. For RayStation 4.7, the issue applies also to machines with fixed jaws, regardless of MLC/jaw position. RayStation is a software system designed for…
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Hazard
- Affected units
- 245
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (8)
- Build numbers 3.5.0.16
- 3.5.1.6
- 4.0.0.14
- 4.0.1.4
- 4.0.2.9
- 4.0.3.4
- 4.5.0.19
- 4.5.1.14 and 4.7.0.15
Distribution
Distributed nationwide across the United States.
Related recalls
Same manufacturer · RAYSEARCH LABORATORIES AB
See all →- HighRayCare 2024A SP1 Oncology Information System Software Recall
FDA (Devices) · 2026-07-08
- HighRayCare 2024A SP4 Oncology System Beam Set Delivery Note Data Loss
FDA (Devices) · 2026-07-08
- HighRayStation 2024A treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation 2023B and 2024A recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
- HighRayStation v2025 treatment planning recalled over dose that fails to invalidate
FDA (Devices) · 2026-01-28
Same hazard · dose calculation error
See all →- HighRadiation Therapy Planning System Density Calculation Inconsistency in RayStation
FDA (Devices) · 2025-02-26
- HighRayStation radiation therapy planning software density uncertainty calculation inconsistency
FDA (Devices) · 2025-02-26
- HighMedical device recall: RayStation radiation therapy software dose calculation inconsistency
FDA (Devices) · 2025-02-26
- HighRayStation Radiation Therapy Planning Software Recall for Density Calculation Inconsistency
FDA (Devices) · 2025-02-26
- HighRadiation Therapy Planning Software Correction for Density Uncertainty in Dose Calculations
FDA (Devices) · 2025-02-26
Similar recalls
- HighRayStation Radiation Therapy Software Recalled for Dose Calculation Errors
FDA (Devices) · 2016-12-14
- HighRayStation Radiation Therapy Software Recalled for Dose Calculation Accuracy Issues
FDA (Devices) · 2018-07-04
- HighRaySearch RayStation Software Recalled for Dose Calculation Errors
FDA (Devices) · 2015-06-03
- HighRayStation Treatment Planning Software Recalled for Dose Calculation Errors
FDA (Devices) · 2020-02-05
- HighRaySearch Recalls Radiation Therapy Planning Software Due to Dose Calculation Errors
FDA (Devices) · 2019-07-10
- HighRayStation Radiation Therapy Software Recalled for Potential Dose Errors
FDA (Devices) · 2014-04-09