Siemens HSV-2 Test Kits Recalled for Potential False Positive Results
Siemens is recalling ADVIA Centaur4P Herpes-2 IgG test kits due to stability and calibration failures that may cause false positive results.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a risk of harm (false positive diagnostic results) where no specific injuries or illnesses have been reported, fitting the 'High' severity criteria for risk-of-harm products.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling ADVIA Centaur4P Herpes-2 IgG (HSV2) test kits. The recall involves 390 kits total, with 366 distributed domestically in the United States and 24 distributed in foreign countries including Nepal, Chile, Georgia, and Spain. The affected lots are B33255, B33269, and B33413.
The recall was initiated to correct Onboard Stability and Calibration Frequency (Interval) Failures on the ADVIA Centaur XP System and ADVIA Centaur XPT System. These technical failures may result in potential false positive results for Herpes Simplex Virus type 2 (HSV-2).
Healthcare facilities and laboratories that have received these kits should stop using them and contact Siemens Healthcare Diagnostics for further instructions. Consumers who have received test results from these specific kits should consult their healthcare provider to discuss the potential for false positives.
The recalled product
- Product
- Enzyme linked immunosorbent assay, for herpes simplex virus, hsv-2
- Manufacturer
- Siemens Healthcare Diagnostics, Inc
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- ***Updated as of 3/25/2021*** Lot #: B33255
- B33269
- B33413. UDI #: (01)00630414016726(10)B33255(17)20210625
- (01)00630414016726(10)B33269(17)20210629
- (01)00630414016726(10)B33413(17)20210819. Updated as of 3/25/2021
Distribution
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · false positive
See all →- HighAlere San Diego Recalls hCG Urine Pregnancy Tests for False Positives
FDA (Devices) · 2026-10-07
- HighHealthmark ProChek-II Medical Tests Recalled for Expired Ingredient
FDA (Devices) · 2026-09-02
- HighAmyloid 1-42-N plasma cartridges recalled over falsely elevated results
FDA (Devices) · 2026-02-11
- HighpTau 217 plasma cartridges recalled over falsely elevated ratio results
FDA (Devices) · 2026-02-11
- HighAmyloid plasma controls recalled over falsely elevated ratio results
FDA (Devices) · 2026-02-11
Similar recalls
- HighHepatitis B Diagnostic Test Kits Recalled for Test Interference
FDA (Devices) · 2023-03-01
- HighSiemens ADVIA Centaur HBc Total Kit Recalled for Increased Imprecision
FDA (Devices) · 2014-10-29
- ModerateSiemens ADVIA Centaur aTG Test Kits Recalled for Positive Bias
FDA (Devices) · 2019-05-08
- HighSiemens Recalls HER-2/neu Assay Kits Due to Inaccurate Results
FDA (Devices) · 2020-10-28
- ModerateSiemens ADVIA Centaur CA 125II Assay Recalled for Over-Recovery
FDA (Devices) · 2018-05-02
- HighADVIA Centaur SARS-CoV-2 Antigen Assay Recalled for Insufficient Validation Data
FDA (Devices) · 2021-11-10