Dako Automated Slide Stainers Recalled for Potential Diagnostic Errors

Dako North America Inc. is recalling four models of automated slide stainers because a defect may cause buffer leakage, leading to false negative diagnostic results.

What this means for you

Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.

Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a risk of harm (false negative diagnostic results) where injury has not yet been reported, fitting the 'High' severity criteria.

Plain-English summary

Dako North America Inc. is recalling 409 units of four models of automated slide stainers: Autostainer Link 48 (AS480), Autostainer Plus Link Instrument (AS100), Autostainer (S3400), and Autostainer Plus (S3800). These devices are used for in vitro diagnostic purposes. The recall affects units manufactured between December 2013 and August 2014.

The recall is due to a defect in the syringe assembly stopcock component. In some circumstances, this defect may cause excess buffer to leak onto specific slide locations: location 34 on the AS480, or locations 35 and 36 on the AS100, S3400, and S3800. This leakage can result in false negative test results, which may affect patient diagnosis.

The affected devices were distributed worldwide, including the United States and several countries in Europe, Asia, and Oceania. Each device is labeled with a unique serial number. Users should check their device serial numbers against the recall information and contact Dako North America Inc. for further instructions.

The recalled product

Product
Autostainer Link 48 (AS480), Autostainer Plus Link Instrument (AS100), Autostainer (S3400), and Autostainer Plus (S3800). Automated slide stainer for in vitro diagnostic use.
Affected units
409

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (3)

  • 992640 Rev. 02
  • 992085 Rev.no H
  • 992625 Rev.no F

Distribution

Distributed nationwide across the United States.