Scleral IOL markers recalled over voids in sterile pouch seals
Vortex Surgical is recalling 14,789 units because there may be voids in the seal of the Tyvek pouches, which compromises the sterile barrier.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: Scored 3 under the risk-of-harm criterion: the firm states a compromised sterile barrier could lead to infection, and no injuries are reported.
Plain-English summary
This recall by Vortex Surgical Inc. covers the Vortex Surgical Lindsell Sutured IOL Marker, VS0390 Scleral Marker.
The record states there may be voids located in the seal of Tyvek pouches associated with the identified catalog numbers and lots, and that a compromised sterile barrier has the potential of bioburden contamination which could lead to infection.
The record lists 14,789 units, of which 8,651 are in the US and 6,138 outside it, with pouch and box UDI codes and lot numbers recorded in the enforcement record. Distribution was worldwide, including nationwide in the United States across the states listed in the record. The FDA has classified the action as Class II.
The enforcement record reports no injuries. Facilities should quarantine affected stock and follow Vortex Surgical's instructions.
The recalled product
- Product
- Vortex Surgical Lindsell Sutured IOL Marker, VS0390 - Scleral Marker
- Manufacturer
- Vortex Surgical Inc.
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Pouch UDI 810123483655 Box UDI 810123483662 Lot 2408035
Distribution
Part of a larger recall action
This product is one of 13 recalled by Vortex Surgical Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 13 products in this recall →Related recalls
Same manufacturer · Vortex Surgical Inc.
See all →- HighInjection kits recalled over voids in sterile Tyvek pouch seals
FDA (Devices) · 2026-02-11
- HighBackflush instruments recalled over voids in sterile pouch seals
FDA (Devices) · 2026-02-11
- HighOertli laser probes recalled over voids in sterile pouch seals
FDA (Devices) · 2026-02-11
- HighSubretinal injection cannulas recalled over voids in pouch seals
FDA (Devices) · 2026-02-11
- HighSurgical laser probes recalled over voids in sterile pouch seals
FDA (Devices) · 2026-02-11
Same hazard · sterile barrier
See all →- HighSplashwire hydrophilic guide wire pouches have unsealed portions
FDA (Devices) · 2026-06-24
- HighCovidien EndoStitch loading unit pouches may have sterility-breaking holes
FDA (Devices) · 2026-06-10
- HighReprocessed electrophysiology catheters recalled over incomplete sterile seals
FDA (Devices) · 2026-05-20
- HighMediHoney wound gel recalled over potential packaging failures
FDA (Devices) · 2026-02-25
- HighCatheter cutdown trays recalled over damage to the outer tray
FDA (Devices) · 2026-02-25
Similar recalls
- HighSurgical Product Recalled Due to Torn Pouches Compromising Sterile Barrier
FDA (Devices) · 2024-07-24
- HighVial Decanter Medical Device Recalled for Compromised Sterile Packaging
FDA (Devices) · 2024-08-28
- HighOtopore Square Outer Ear Wound Dressing Recall for Sterility Seal Defect
FDA (Devices) · 2025-04-09
- HighVersaCross access kits may have a hole in the sterile pouch
FDA (Devices) · 2025-07-16
- HighIrriMAX Recalls Irrisept Wound Debridement System Due to Seal Defect
FDA (Devices) · 2019-10-30
- HighSurgilon Braided Nylon sutures recalled due to sterile barrier defect
FDA (Devices) · 2023-03-08