The Recall Desk
ModerateFDA (Devices)·Z-1235-2013·Announced 2013-05-15

Stryker Irrigator Tips Recalled for Incorrect Sterilization Temperature Instructions

Stryker Endoscopy is recalling reusable irrigator tips because the instructions listed incorrect pre-vacuum steam sterilization temperatures.

What this means for you

Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.

Our severity reasoning: The recall is classified as FDA Class II, but the source text identifies the issue as a labeling error regarding sterilization temperatures without reporting any specific illnesses or injuries, fitting the criteria for a moderate severity voluntary precautionary recall.

Plain-English summary

Stryker Endoscopy is recalling 147,921 units of Stryker Irrigator Reusable Tips (Manufacturer Part Number 250-070-489). These tips are used to clean and clear the operative site during laparoscopic surgery. The recall covers all sizes of the 5mm x 32cm Regular Tip with No Holes, Non-Reflective Finish.

The recall was initiated because the Instructions for Use incorrectly listed the pre-vacuum steam sterilization (wrapped method) temperature as 121-123°C, instead of the correct 132-133°C. This labeling error could result in improper sterilization of the devices.

The affected products were distributed worldwide, including the United States (nationwide), Mexico, Portugal, Latin America, India, Greece, Taiwan, Singapore, New Zealand, the Philippines, the Netherlands, France, Australia, Canada, Japan, Scandinavia, Iberia, Brazil, Romania, South Africa, Chile, Korea, and China. Users should stop using the affected tips and contact Stryker Endoscopy for instructions on how to proceed.

The recalled product

Product
Stryker Irrigator Reusable Tips: 5mm X 32cm Regular Tip with No Holes, Non-Reflective Finish. Suction irrigator tips aid in cleaning and clearing the operative site during laparoscopic surgery.
Manufacturer
Stryker Endoscopy
Affected units
147,921

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (1)

  • Manufacturer Part Number: 250-070-489

Distribution

Distributed nationwide across the United States.

Part of a larger recall action

This product is one of 20 recalled by Stryker Endoscopy under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.

See all 20 products in this recall →