Siemens Dimension Enzymatic Creatinine Reagent Recalled for Interference
Siemens Healthcare Diagnostics is recalling Dimension Enzymatic Creatinine reagent because it may produce falsely low results for patients taking phenindione.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk-of-harm product (medical device reagent) where injury has not yet been reported, fitting the High severity criteria.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling 2,198 units of Dimension Enzymatic Creatinine (EZCR) reagent, SMN 10471520. The recall covers all lots and was distributed nationwide in the United States.
The recall is due to a potential for falsely depressed creatinine results when the enzymatic methodology is used for patients on phenindione therapy. The source notes that this interference has not been observed with the creatinine Jaffe methodology.
Healthcare facilities should stop using the affected reagent for patients on phenindione therapy. Consumers should contact their healthcare provider if they have had recent creatinine tests using this specific reagent while taking phenindione.
The recalled product
- Product
- Dimension Enzymatic Creatinine (EZCR) reagent, SMN 10471520
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Affected units
- 2,198
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- UDI Number: 00842768030512 Lots: All lots
Distribution
Distributed nationwide across the United States.
Part of a larger recall action
This product is one of 4 recalled by Siemens Healthcare Diagnostics, Inc. under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 4 products in this recall →Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- HighSiemens Atellica CH C-Reactive Protein Reagent Recalled for Falsely Elevated Results
FDA (Devices) · 2026-10-07
- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · test interference
See all →- HighSiemens Atellica Lab Analyzers Risk Falsely Elevated Cholesterol Results
FDA (Devices) · 2024-04-24
- HighFetal Membrane Rupture Test Kits Recalled for False Positive Results
FDA (Devices) · 2024-04-24
- HighAbbott Alinity i Anti-HCV Reagent Kit Recall for Falsely Elevated Test Results
FDA (Devices) · 2024-04-10
- HighHepatitis B Diagnostic Test Kits Recalled for Test Interference
FDA (Devices) · 2023-03-01
- HighFDA Recalls V8 Immunodisplacement Kit Due to Microbial Contamination
FDA (Devices) · 2022-06-15
Similar recalls
- HighSiemens Dimension Enzymatic Creatinine Recalled for Acetaminophen Interference
FDA (Devices) · 2020-06-17
- HighSiemens Dimension Enzymatic Creatinine Assay Recalled for Interference
FDA (Devices) · 2016-05-18
- HighSiemens Dimension Vista Enzymatic Creatinine Recalled for Acetaminophen Interference
FDA (Devices) · 2020-06-17
- HighSiemens Dimension Vista Enzymatic Creatine Assays Recalled for Interference
FDA (Devices) · 2016-05-18
- ModerateDimension Creatinine Flex Reagent Cartridge Recalled for Imprecision Results
FDA (Devices) · 2026-05-06
- ModerateSiemens Creatine Kinase MB Assay Recalled for Drug Interference
FDA (Devices) · 2018-05-16