Brainlab RT Elements Software Recalled for Incorrect Dose Calculation
Brainlab AG is recalling specific versions of RT Elements software due to a potential for incorrect dose distribution calculations when using the Pencil Beam algorithm on the GPU.
What this means for you
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device software defect that could lead to incorrect radiation dose calculations, posing a risk of harm to patients, although no specific injuries or illnesses are reported in the source text.
Plain-English summary
Brainlab AG is recalling specific versions of its RT Elements software applications, including Cranial SRS, Spine SRS, Multiple Brain Mets SRS, and RT QA. The affected software versions are 1.0.0 and 1.5.0 for Cranial SRS, Spine SRS, and RT QA, and version 1.5.0 for Multiple Brain Mets SRS.
The recall is issued because there is a potential for an incorrect dose distribution calculation under specific circumstances. This issue occurs when using the Pencil Beam algorithm on the GPU (graphics card), which is the default system setting for these applications. The software is used for radiation treatment planning for stereotactic, conformal, and computer-planned Linac-based radiation treatment of cranial, head and neck, and extracranial lesions.
The recall affects 164 units in the United States and 333 units outside the US. Distribution was worldwide, including nationwide distribution in the US across multiple states and various international locations. Users of the affected software should contact Brainlab AG for further instructions on corrective actions.
The recalled product
- Product
- RT Elements Software revisions of the RT Elements applications have a specific software version number. Specifically the following RT Elements applications/versions are affected: - Cranial SRS 1.0.0 and 1.5.0 - Spine SRS 1.0.0 and 1.5.0 - Multiple Brain Mets SRS 1.5.0 -
- Manufacturer
- Brainlab AG
Distribution
Related recalls
Same manufacturer · Brainlab AG
See all →- HighMedical Device Software May Merge Patient Records Without Notification
FDA (Devices) · 2025-03-26
- HighSurgical navigation system recalled for potential mechanical instability
FDA (Devices) · 2023-04-05
- HighRadiation Therapy Software May Not Properly Execute Heart-Protection Breath Hold
FDA (Devices) · 2023-03-22
- HighExacTrac Dynamic Software Yaw Angle Error May Affect Radiotherapy Positioning
FDA (Devices) · 2022-03-16
- HighRadiotherapy positioning software may delay patient movement display
FDA (Devices) · 2021-06-16
Same hazard · incorrect dose calculation
See all →- HighRayStation Radiation Therapy Software Recalled for Incorrect Dose Calculation
FDA (Devices) · 2023-07-05
- HighRayStation software recall due to incorrect total dose display
FDA (Devices) · 2021-05-26
- HighRayStation Treatment Planning Software Recalled for Dose Calculation Errors
FDA (Devices) · 2020-02-05
- HighElekta Monaco Radiation Treatment Planning System Software Recall
FDA (Devices) · 2019-12-18
- HighElekta Monaco RTP System Recalled for Potential Dose Calculation Errors
FDA (Devices) · 2019-10-02