Siemens CentraLink Software Recalled for Potential Sample Type Omission
Siemens is recalling CentraLink Data Management System software versions v13x, v14x, and v15x because the system may download orders to the ADVIA Automation System without specifying the sample type.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, and the source describes a 'remote possibility' of a data management error without reporting any illnesses, injuries, or hospitalizations, fitting the criteria for a moderate severity voluntary precautionary recall.
Plain-English summary
Siemens Healthcare Diagnostics, Inc. is recalling the CentraLink Data Management System Software, specifically versions v13x, v14x, and v15x. The CentraLink system is a network solution provider and multi-system data manager used to consolidate data from laboratory instruments and lab automation systems (LAS). It allows operators to review and edit patient and quality control results from a single location.
The recall is due to a remote possibility that the CentraLink software may download an order to the ADVIA Automation System without specifying the sample type. This issue can occur when an order is received from the Laboratory Information System (LIS) without a sample type, requiring the sample type to be set in CentraLink based on the test in the order.
The recall affects 3,893 units distributed worldwide, including nationwide distribution in the United States to the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, NC, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, and WV, as well as to Canada and other countries. Affected product codes are listed in the source metadata. Users of this software should contact Siemens Healthcare Diagnostics for further instructions on corrective actions.
The recalled product
- Product
- CentraLink¿ Data Management System Software Versions: v13x,v14x,v15x The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Affected units
- 3,893
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (20)
- Product codes: 10485061
- 10484212
- 10485064
- 10483439
- 10471393
- 10484552
- 10814296
- 10815474
- 11220480
- 10481936
- 10481937
- 10481938
- 10482170
- 10482936
- 10483437
- 10483438
- 10483440
- 10484211
- 10484213
- 10485060
Distribution
Related recalls
Same manufacturer · Siemens Healthcare Diagnostics, Inc.
See all →- SevereSiemens Recalls Emit 2000 Quinidine Assay Due to Precision Issues
FDA (Devices) · 2026-08-26
- SevereSiemens Quinidine Assay Recalled for Precision Errors
FDA (Devices) · 2026-08-26
- HighSiemens Atellica CH Urine Albumin In Vitro Diagnostic Recall
FDA (Devices) · 2026-08-12
- ModerateSiemens Tricyclic Antidepressant Calibrator/Controls Recall Lot U2
FDA (Devices) · 2026-08-12
- LowSyva Emit Disopyramide Enzyme Immunoassay Calibrator Recalled for Sodium Azide Concentration
FDA (Devices) · 2026-08-12
Same hazard · data integrity
See all →- ModerateVentana Recalls navify Digital Pathology Software Due to Image Metadata Discrepancies
FDA (Devices) · 2026-08-19
- SevereGE Healthcare MAC VU360 Electrocardiograph Software Recall
FDA (Devices) · 2026-08-12
- SevereCopan WASPLab Software Recall Over Results Overwrite Issue
FDA (Devices) · 2026-08-12
- SevereMedline spinal anesthesia trays recalled over Huons bupivacaine quality concerns
FDA (Devices) · 2026-06-03
- SevereMedline pain and Whitacre spinal trays recalled over bupivacaine concerns
FDA (Devices) · 2026-06-03