The Recall Desk
Agency classClass IIOur severity:SevereFDA (Devices)·Z-1201-2014·Announced 2014-03-19

Mar Cor Purification Reverse Osmosis Systems Recalled for Fire Risk

Mar Cor Purification is recalling 130 CWP Reverse Osmosis Systems (Models 101, 102, 104, and 106) due to a potential fire hazard from resistive heating in the pump head assembly.

The agency’s classification

FDA (Devices): Class II

Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.

This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work

What this means for you

Our assessment, not the agency’s: severity 4/5 (Severe).

The hazard poses a serious risk of illness, injury, or structural failure. Stop using the product immediately and contact the manufacturer for a refund or repair.

Our severity reasoning: The recall involves a potential fire hazard (smoke/fire) which constitutes a significant injury/property-damage risk, aligning with the 'Severe' criteria for risk-of-harm products.

Plain-English summary

Mar Cor Purification is recalling 130 units of the CWP Reverse Osmosis System, specifically Models 101, 102, 104, and 106. The recall is due to a potential defect in the crimp connection at the pump head assembly. The reverse osmosis pump head wiring could build resistance, resulting in resistive heating of the surrounding components.

If this heating occurs, the plastic components of the pump head assembly may melt, potentially creating smoke or fire and causing the unit to shut down. The affected units were distributed nationwide in the United States, with no distribution outside the US. Specific serial numbers for the affected units are listed in the official recall documentation.

Consumers who own these specific models should stop using the product immediately. They should contact Mar Cor Purification for instructions on how to obtain a refund or replacement, as per standard recall procedures for this agency classification.

The recalled product

Product
Mar Cor Purification CWP Reverse Osmosis System. Models 101, 102, 104, and 106.
Affected units
130

Is your product affected?

Check your packaging against the codes below. If any of them match, the product is part of this recall.

Lot numbers (20)

  • serial numbers: 1289064
  • 1289069
  • 1289159
  • 1289259
  • 1289387
  • 1289650
  • 1289747
  • 1289826
  • 1289949
  • 1290088
  • 1290151
  • 1290333
  • 1290367
  • 1290480
  • 1290489
  • 1290845
  • 1290988
  • 1291042
  • 1291120
  • 1291233

Distribution

Distributed nationwide across the United States.

Same manufacturer · Mar Cor Purification