Acumed Recalls 6-Hole Pubic Symphysis Plates Due to Material Error
Acumed LLC is recalling 15 units of 6-hole Pubic Symphysis plates because they were manufactured with the wrong titanium alloy.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a medical device with a material specification error (wrong alloy) where no injuries or illnesses have been reported, fitting the rubric for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Acumed LLC is recalling 15 units of the 6-hole Pubic Symphysis plate (PN 70-0451), lot number 383813. These pelvic bone plates are internal fixation devices used by surgeons for fractures, fusions, and osteotomies of the pelvic ring.
The recall was initiated because the recalled batch was manufactured using Titanium Alloy (Ti-6Al-4V ELI per ASTM F135) instead of the specified Commercially Pure Titanium (ASTM F67 - CP Grade 4). The source text does not report any illnesses, injuries, or adverse events associated with this material discrepancy.
The affected units were distributed worldwide to Costa Rica, Malaysia, Mexico, South Africa, and Spain. Healthcare facilities and patients who have received these specific plates should contact Acumed LLC for instructions on removal or replacement.
The recalled product
- Product
- 6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varying lengths that function as internal fixation devices for fractures, fusions and osteotomies of the pelvic ring. The plates are strategically pre-contoured…
- Manufacturer
- Acumed LLC
- Affected units
- 15
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Lot number: 383813
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Acumed LLC
See all →- HighAcumed Hexalob Screws Recalled for Risk of Fracture During Use
FDA (Devices) · 2026-07-01
- HighBone plates and surgical screws distributed without FDA pre-market clearance
FDA (Devices) · 2024-03-06
- HighAcumed RibLoc U Plus 90 Surgical Guide Recalled for Potential Breakage
FDA (Devices) · 2024-01-03
- HighDual-Trak Clavicle Screws recalled: distributed without FDA pre-market clearance
FDA (Devices) · 2023-01-04
- ModerateAcumed Acutrak 2 Bone Screws Recalled for Distribution Without FDA Clearance
FDA (Devices) · 2023-01-04
Same hazard · material mis specification
See all →- ModerateSpecialty Appliance Works Recalls Universal Nuts for Incorrect Material Composition
FDA (Devices) · 2015-04-29
- ModerateSpecialty Appliance Works Recalls Orthodontic Universal Nuts Due to Material Error
FDA (Devices) · 2015-04-29
Similar recalls
- LowApex Knee System Tibial Baseplate Recalled for Packaging Date Error
FDA (Devices) · 2018-03-28
- ModerateBiomet Recalls Proximal Humerus Plating System Due to Incomplete Design Files
FDA (Devices) · 2020-06-17
- ModerateBiomet A.L.P.S. Proximal Humerus Plating System Recalled for Design History Deficiencies
FDA (Devices) · 2020-06-17
- ModerateBiomet A.L.P.S. Proximal Humerus Plating System Recalled for Design Documentation Deficiencies
FDA (Devices) · 2020-06-17
- ModerateBiomet Recalls Proximal Humerus Plate Templates Due to Incomplete Design Files
FDA (Devices) · 2020-06-17
- ModerateBiomet Recalls Proximal Humerus Plating System Due to Incomplete Design History
FDA (Devices) · 2020-06-17