Roche Recalls cobas MRSA/SA Nucleic Acid Tests for Decreased Performance
Roche Molecular Systems is recalling cobas MRSA/SA tests worldwide due to decreased performance that may lead to screening failures and potential infection spread.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II and involves a risk-of-harm product (infection spread) where injury has not yet been reported, fitting the 'High' severity criteria.
Plain-English summary
Roche Molecular Systems, Inc. is recalling 136 kits of the cobas MRSA/SA nucleic acid test intended for screening for MRSA/SA. Each kit contains 20 tubes. The recall is classified as Class II by the U.S. Food and Drug Administration.
The recall is being issued because the tests may have decreased performance compared to the analytical sensitivity for MRSA detection. A screening failure can lead to potential infection in the colonized patient being tested as well as the spread of MRSA to others.
The affected lots are 70712A (6 kits), 70601W (76 kits), 70713A (4 kits), 70718A (50 kits), and 70809A (kits were not distributed to end users). The product was distributed worldwide to Austria, Denmark, Germany, Singapore, the Netherlands, Hong Kong, and Switzerland. The United States is not impacted, as the cobas MRSA/SA nucleic acid test for use on the cobas liat system is not available in the US.
The recalled product
- Product
- Nucleic Acid test intended for screening for MRSA/SA. Each kit has 20 tubes. in vitro diagnostic.
- Manufacturer
- Roche Molecular Systems, Inc.
- Affected units
- 136
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (5)
- Lot #s: 70712A (6 kits)
- 70601W (76 kits)
- 70713A (4 kits)
- 70718A (50 kits)
- and 70809A (kits were not distributed to end users).
Distribution
Distribution scope not specified by the agency.
Related recalls
Same manufacturer · Roche Molecular Systems, Inc.
See all →- HighRoche cobas Influenza Test Recall Due to False Negative Results
FDA (Devices) · 2023-02-15
- HighRoche Influenza A/B and RSV Test Kit Recalled for False Negative Results
FDA (Devices) · 2023-02-15
- HighRoche cobas Flu/COVID assay recalled for false negative influenza results
FDA (Devices) · 2023-02-15
- HighMedical Device Recall: Roche cobas 5800 False Results Due to Baseline Anomalies
FDA (Devices) · 2022-10-26
- ModerateCobas Liat respiratory virus analyzer units recalled for false Influenza B results
FDA (Devices) · 2022-04-06
Same hazard · infection
See all →- SevereSafeguard US Operating LLC Recalls Compression Bandages Due to Degradation
FDA (Devices) · 2026-10-14
- SevereStryker Bard EP XT Decapolar Steerable Diagnostic EP Catheter Recalled for Sterile Barrier Breach
FDA (Devices) · 2026-10-14
- HighEmbosphere Microspheres Recalled Due to Packaging Sterility Concerns
FDA (Devices) · 2026-10-07
- SevereAurolab Trust Recalls FOCUS PI Eye Drops Due to Sterility Concerns
FDA (Drugs) · 2026-10-07
- HighFDA Recalls Performance Pro Nitrile Gloves Due to Defects
FDA (Devices) · 2026-10-07
Similar recalls
- SevereRoche cobas Liat Analyzer Respiratory Virus Panel Recall
FDA (Devices) · 2021-04-21
- HighRoche cobas Influenza Test Recall Due to False Negative Results
FDA (Devices) · 2023-02-15
- HighRoche Recalls cobas SARS-CoV-2 and Influenza Assay Reagents
FDA (Devices) · 2021-04-21
- HighRoche Influenza A/B and RSV Test Kit Recalled for False Negative Results
FDA (Devices) · 2023-02-15
- Highcobas SARS-CoV-2 and Influenza A/B Test Recalled Due to False Positive Results
FDA (Devices) · 2021-09-29
- HighRoche cobas Flu/COVID assay recalled for false negative influenza results
FDA (Devices) · 2023-02-15