Toshiba DRAD-3000E Wireless System Recalled for Image Transmission Failure
Toshiba is recalling 28 DRAD-3000E FPD Wireless Systems because they fail to complete image transmission, resulting in no radiographic image.
What this means for you
Minor labeling or packaging issue, or a voluntary precautionary recall. Risk of harm is low but real — follow the manufacturer's instructions.
Our severity reasoning: The recall is classified as FDA Class II, but the source text describes a functional software error (image transmission failure) with no reported injuries, illnesses, or physical hazards, fitting the criteria for a moderate severity issue.
Plain-English summary
Toshiba American Medical Systems Inc is recalling 28 units of the DRAD-3000E FPD Wireless System. This medical device is intended for use with ceiling-suspended tube supports and flat panel detectors for radiography of the head, chest, abdomen, spine, neck, and limbs.
The recall was initiated because users reported that when performing radiography, a message window appeared on the monitor stating that imaging transmission was not completed and no radiographic image was produced. Users who selected the "OK" button to re-acquire the image data encountered the same error repeatedly. The issue was resolved only when the user selected "Cancel" to terminate the re-acquisition process.
The affected units were distributed nationwide to Pennsylvania, Utah, Florida, New York, Louisiana, West Virginia, New Jersey, Ohio, Wisconsin, and Texas. The specific serial number for the affected unit is USB1522023. Users should contact Toshiba American Medical Systems Inc for further instructions on how to address this defect.
The recalled product
- Product
- Toshiba DRAD-3000E FPD Wireless System Product Usage The DRAD-3000¿ is intended for use with the ceiling-suspended tube support, high voltage generator, and bucky stand or bucky table incorporating a fixed or detachable (portable) flat panel detector for radiography of the h
- Manufacturer
- Toshiba American Medical Systems Inc
- Affected units
- 28
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Serial number : USB1522023
Distribution
Related recalls
Same manufacturer · Toshiba American Medical Systems Inc
See all →- ModerateToshiba Diagnostic Ultrasound Systems Recalled for Power Supply Defect
FDA (Devices) · 2018-07-04
- ModerateToshiba INFX-8000H Fluoroscopic X-Ray Systems Recalled for Ceiling Drive Defect
FDA (Devices) · 2018-03-14
- HighToshiba Fluoroscopic X-Ray Systems Recalled for Ceiling Drive Bolt Loosening
FDA (Devices) · 2018-03-14
- HighToshiba INFX-8000F Fluoroscopic X-Ray Systems Recalled for Ceiling Drive Defect
FDA (Devices) · 2018-03-14
- HighToshiba INFX-8000C Fluoroscopic X-Ray Systems Recalled for Ceiling Drive Defect
FDA (Devices) · 2018-03-14
Same hazard · software error
See all →- HighQuantum Rehab R-TRAK Powered Wheelchair Controller Recalled Due to Software Issue
FDA (Devices) · 2026-08-12
- HighDASH Rapid PCR Instrument Recalled Due to Software Issue Causing False Co-infection Results
FDA (Devices) · 2026-08-12
- HighQuantum Outback Powered Wheelchair Controller Recalled Due to Unexpected Braking Risk
FDA (Devices) · 2026-08-12
- HighREMISOL Advance Software Version 2.5 Recalled for Incorrect Test Result Reporting
FDA (Devices) · 2026-08-12
- HighQuantum Rehab Q6 Edge 3 Powered Wheelchair Controller Recalled for Unexpected Braking
FDA (Devices) · 2026-08-12