Siemens AXIOM Artis angiographic x-ray systems recalled for coolant leaks
Siemens is recalling 22 AXIOM Artis angiographic x-ray systems because leaking coolant couplings can damage the flat panel detector, potentially requiring treatment cancellation or patient transfer.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall is classified as FDA Class II. The hazard involves a risk of harm (treatment interruption or patient transfer) due to equipment failure, but no injuries or illnesses are reported in the source text, fitting the 'High' severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Siemens Medical Solutions USA, Inc. is recalling 22 units of the AXIOM Artis angiographic x-ray system, specifically model numbers 7412807, 7413078, and 7728392. The affected units are distributed in the United States across the states of Arizona, California, Florida, Iowa, Illinois, Louisiana, Michigan, North Carolina, New Jersey, Ohio, Oklahoma, and Texas.
The recall is due to a potential defect where plastic coolant couplings may leak. If this occurs, liquid can enter the flat panel detector, leading to a failure of the detector. This malfunction may necessitate the cancellation or restart of a medical treatment, or the transfer of the patient to an alternate or another system.
Healthcare facilities should identify affected units by checking the serial numbers listed in the official recall notice. Siemens is the recalling firm and is responsible for addressing the defect. No specific instructions for end-users are provided in the source text, as the product is a medical device used in clinical settings.
The recalled product
- Product
- AXIOM Artis angiographic x-ray system, Model Numbers: 7412807, 7413078, 7728392 AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese…
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Affected units
- 22
Distribution
Related recalls
Same manufacturer · Siemens Medical Solutions USA, Inc
See all →- ModerateARTIS icono biplane medical imaging system detector lift malfunction
FDA (Devices) · 2026-07-08
- HighARTIS icono floor Medical Imaging System Detector Movement Loss Recall
FDA (Devices) · 2026-07-08
- ModerateARTIS icono ceiling medical imaging monitor suspension system
FDA (Devices) · 2026-07-08
- HighARTIS Icono Ceiling Model 11328100 Loss of Movement Capability Recall
FDA (Devices) · 2026-07-08
- HighARTIS pheno systems may display zero dose while dose is applied
FDA (Devices) · 2026-06-10
Same hazard · equipment failure
See all →- High
- HighHeidelberg SPECTRALIS Flex Module Recall Due to Wheel Detachment Risk
FDA (Devices) · 2026-09-30
- HighMobile Digital Radiographic System Arm Drop Risk During Patient Exams
FDA (Devices) · 2026-07-22
- HighGE Optima XR200/220/240amx Radiographic System Column Safety Mechanism
FDA (Devices) · 2026-07-22
- SevereOmnipod Eros Pod Insulin Delivery Failure Risk
FDA (Devices) · 2026-07-08
Similar recalls
- HighSiemens AXIOM Artis X-Ray Systems Recalled for Cooling System Leak
FDA (Devices) · 2016-04-06
- HighSiemens AXIOM Artis Angiography System Recalled for Defective Sealing
FDA (Devices) · 2018-08-01
- HighSiemens AXIOM Artis zee and Q/Q.zen X-ray Systems Recalled for Cooling Leak
FDA (Devices) · 2017-09-06
- HighSiemens AXIOM Artis zee/zeego X-Ray Systems Recalled for Cooling Leak
FDA (Devices) · 2016-04-06
- SevereSiemens AXIOM Artis Systems Recalled for Potential Gantry Drop Risk
FDA (Devices) · 2014-04-02
- ModerateSiemens AXIOM Artis Angiography Systems Recalled for Cooling System Contamination
FDA (Devices) · 2016-08-17