Zimmer Metasul Femoral Head Recalled for Potential Sterile Tray Compromise
Zimmer GmbH is recalling 16 Metasul Femoral Heads due to potential sterile tray compromise, which could lead to periprosthetic infection if undetected.
The agency’s classification
FDA (Devices): Class II
Class II is a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences, or where the probability of serious adverse health consequences is remote.
This is the agency’s own classification, taken from its notice. The severity score below is our assessment, not the agency’s. How recall classes work
What this means for you
Our assessment, not the agency’s: severity 3/5 (High).
Real risk of harm even if no illness or injury has been reported yet. Stop using the product and follow the agency's guidance.
Our severity reasoning: The recall involves a risk of periprosthetic infection, a significant health hazard, but the source text describes the risk as 'possible' and does not report any actual illnesses or injuries, fitting the High severity criteria for risk-of-harm products where injury has not yet been reported.
Plain-English summary
Zimmer GmbH is recalling 16 units of the Metasul Femoral Head, a medical device indicated for total hip arthroplasty for the treatment of non-inflammatory degenerative joint disease. The recall is due to a potential compromise of the inner sterile tray.
The affected units are identified by item number 734028800 and lot number 61487614. These devices were distributed domestically to 45 states and internationally to Canada and Japan. If a compromised tray is detected, it may result in a delay in surgery. If an undetected breach occurs in both the inner and outer trays and contamination is present, a periprosthetic infection is possible, which could lead to revision surgery.
Healthcare providers and facilities should check their inventory for the specific item and lot numbers listed above. If the product is found, it should be quarantined and not used until further notice.
The recalled product
- Product
- Metasul Femoral Head - indicated for total hip arthroplasty for treatment of patient conditions of non-inflammatory degenerative joint disease.
- Manufacturer
- Zimmer Gmbh
- Affected units
- 16
Is your product affected?
Check your packaging against the codes below. If any of them match, the product is part of this recall.
Lot numbers (1)
- Item numbers: 734028800 Lot #'s: 61487614
Distribution
Part of a larger recall action
This product is one of 5 recalled by Zimmer Gmbh under a single FDA (Devices) recall action. The agency files each product separately, so the size of the action is not visible from this page alone.
See all 5 products in this recall →Related recalls
Same manufacturer · Zimmer Gmbh
See all →- HighBiolox Taper Sleeve Mislabeling: Wrong Component Size in Hip Replacement
FDA (Devices) · 2024-09-18
- ModerateZimmer GmbH Recalls Mislabeled Biolox Option Taper Sleeve for Hip Replacement
FDA (Devices) · 2024-09-18
- HighZimmer Femoral Nail Recall: Risk of Commingled Angle Versions
FDA (Devices) · 2021-06-09
- HighZimmer Natural Nail cephalomedullary device angle mismatch recall
FDA (Devices) · 2021-06-09
- ModerateZimmer Sidus Stem-Free Shoulder Humeral Head Recall for Incorrect Labeling
FDA (Devices) · 2019-10-09
Same hazard · infection
See all →- SevereSafeguard US Operating LLC Recalls Compression Bandages Due to Degradation
FDA (Devices) · 2026-10-14
- SevereStryker Bard EP XT Decapolar Steerable Diagnostic EP Catheter Recalled for Sterile Barrier Breach
FDA (Devices) · 2026-10-14
- HighEmbosphere Microspheres Recalled Due to Packaging Sterility Concerns
FDA (Devices) · 2026-10-07
- SevereAurolab Trust Recalls FOCUS PI Eye Drops Due to Sterility Concerns
FDA (Drugs) · 2026-10-07
- HighFDA Recalls Performance Pro Nitrile Gloves Due to Defects
FDA (Devices) · 2026-10-07
Similar recalls
- ModerateZimmer Biomet Recalls Metasul Head 40 Total Hip Prosthesis Components
FDA (Devices) · 2016-03-02
- ModerateZimmer Biomet Recalls Metasul Femoral Heads Due to Packaging Adhesion
FDA (Devices) · 2019-11-13
- ModerateZimmer Recalls Metasul Head Implants Due to Packaging Adhesion Issue
FDA (Devices) · 2013-11-06
- SevereZimmer Recalls Provisional Femoral Heads Due to Fit Issues
FDA (Devices) · 2012-10-10
- HighZimmer Manufacturing Recalls Unipolar Femoral Heads Due to Sterility Concerns
FDA (Devices) · 2016-03-23
- ModerateZimmer Manufacturing Recalls Sterile Femoral Heads Due to Sealer Calibration Issues
FDA (Devices) · 2016-03-23